User Manual

133
LIN bus trigger and decode
2 Press to select left/right respectively, press to move downward
cycle, press to back; after selecting the trigger data, press or roll
the wheel to adjust, or press the data on the touch screen, or use the virtual
keyboard to modify;
UART trigger configuration menu descriptions:
a) Start bit trigger at the start bit of the measured signal
b) Stop bit trigger at the stop bit of the measured signal, no matter the measured signal
uses 1, 1.5, or 2 stop bits or not, the trigger will occur at the first stop bit.
c) data Trigger at the specified data bit, when measured signal data bits are effective
as five to eight bits, select the data, and select the trigger relationship as =, >, <, , then
select the trigger data, press or roll the wheel to adjust, or press the data on the
touch screen, using the virtual keyboard to modify.
d) 0: datathe measured signal data bits is nine. Only when the ninth bit is zero, then