User guide

A - 5
OMNIScanner User Guide
NEXT
Test all six pair combinations from both directions using OMNI
Remote
Frequency Range: 1 to 300 MHz
Dynamic Range: >100dB
Measurement Accuracy: Calculated per TIA error model
Accuracy: Significantly exceeds Level III requirements
Measurement Resolution: 0.1dB
Return Loss
Measures Return Loss on all four pairs from both directions using
OMNI
Remote
Frequency Range: 1.0 to 300 MHz
Dynamic Range: > -35 dB
Factory Calibrated to ensure measurement port match to reference
100 ohm impedance.
Measurement accuracy: Significantly exceeds Level III requirements
Measurement resolution: 0.1dB
Attenuation
Tests all four pairs using OMNI
Remote
as far end active signal injector.
Frequency Range: 1.0 to 300 MHz
Dynamic Range: > 60 dB
Measurement Accuracy: Significantly exceeds Level III requirements
Measurement resolution: 0.1dB
ELFEXT (Equal Level FEXT)
Tests from both directions using OMNI
Remote
and reports all 24 pair
combinations.
Frequency Range: 1 to 300 MHz
Dynamic Range: >100dB
Measurement Accuracy: Significantly exceeds Level III requirements
Measurement Resolution: 0.1dB
ACR (Attenuation to Crosstalk Ratio)
Computed for 6 pair combinations from both ends at all frequency points
from the respective pair to pair NEXT and Attenuation measurements.
Bandwidth
Computed for the link or channel under test, bandwidth indicates the
highest frequency for which the worst case pair to pair crosstalk (NEXT)
is at least 3dB less than the attenuation.