User guide

8 - 3
OMNIScanner User Guide
1. Do not disconnect the 2 m Launch Cables that are attached to the
OMNI
Fiber
adapters.
2. Remove the middle Launch Cable without disturbing the couplers.
3. Connect the link under test into the open couplers.
When using Method C, the loss effects of all Launch Cables and couplers
are canceled out.
4. Press .
Loopback Autotest
This configuration is used to test single fibers and launch cables. To run a
Loopback Autotest proceed as follows:
1. Do not disconnect the black 2 m Launch Cable, which is attached to the
TX side of the OMNI
Fiber
adapter (it is circled in the graphic below).
2. Do not disconnect the
coupler from the black
Launch Cable (it is circled
in the graphic below).
3. Disconnect the white
Launch Cable that from the
RX side of the OMNI
Fiber
adapter and the coupler.
Leave the two white connectors dangling.
4. Attach the cable to be tested to the RX connector on the OMNI
Scanner
Fiber adapter and the Coupler.
The one extra coupler will add extra loss (which is considered insignificant)
to your measurement.
5. Press .
RX
TX
850 nm
1300/1310 nm

Fixed Connection
Fixed Connection
