7 - 4
Method C
For Method C three launch cables and two connectors are used to set the
reference. All of the launch cables and connectors are canceled. Use this
method when measuring an end to end system where the patch cables need
to be included in the total loss measurement.
1. Connect the first Launch Cable to the fiber adapter that is attached to
OMNI
Scanner
. The black fiber attaches to the TX (transmitter) side and the
white fiber to the RX (receiver) side.
2. Connect the second Launch Cable to the fiber adapter that is attached to
OMNI
Remote
. The black fiber attaches to the RX (receiver) side and the
white fiber to the TX (transmitter) side.
3. Use two Couplers to connect the two black fiber ends with the third
black Launch Cable.
4. Use two Couplers to connect the two white fiber ends with the third
white Launch Cable.
5. Set the Reference.
6. Disconnect the third Launch Cable, located in the center, without disturb-
ing any connections to the couplers or adapters.
Setting the Reference in Loopback Mode
If a single cable needs to be measured, the reference value is set with one
launch cable.
1. Connect the launch cable to the fiber adapter that is attached to OMNI-
Scanner.
The black fiber attached to the TX (transmit) connector and the
white fiber to the RX (receive) connector.
2. Use a coupler to connect the launch cables on the other end.
3. Set the Reference.
RX
TX
850 nm
1300/1310 nm
RX
TX
850 nm
1300/1310 nm
2 m Launch Cable
Couplers
2 m Launch Cable 2 m Launch Cable