Specifications

6
AFC-II
In some of the sensors, only connector section itself can be
replaced. The other sensor should be replaced as a whole of
sensor PCB.
After replacing the 135/240 AMC-II, MMC-II, the focus
adjustment for each magnification is necessary for each
135/240 AMC-II or MMC-II to be used.
(2) 135/240 AMC-II, MMC-II
Explain the position adjustment of AF sensor.
A adjustment of the worm wheel in the MMC Auto Focus
Section(It is not necessary to adjust AMC-II.)
After replacing the carrier, it is necessary to adjust the focus
to the different magnifications depends on the types of the
carrier, i.e. the AMC-II or the MMC-II.
Mechanical adjustment (Film carrier section)
33_6
6030
Explanation
There are 2 types of “mount for adjusting the scanner” and “head height adjustment tool” to be
used for MMC-II/AMC-II adjustment. (currently used type and initial type)
The scanner adjustment mount (currently used type) is available for AMC-II/MMC-II.
The scanner adjustment mount (initial type) is not available for AMC.
 Show both of scanner adjustment mounts.