MS-710 Scanner User's Manual P/N 83-000700 REV I
Copyright © 2000 by Microscan Systems, Inc., 1201 S.W. 7th Street, Renton, Washington, U.S.A. 98055 (425) 226-5700 FAX: (425) 226-8682 ISO 9001/Certification No. US96/0465 All rights reserved. The information contained herein is proprietary and is provided solely for the purpose of allowing customers to operate and/or service Microscan manufactured equipment and is not to be released, reproduced, or used for any other purpose without written permission of Microscan.
Table of Contents Table of Contents ................................................................ iii List of Illustrations ................................................................v List of Tables ...................................................................... vi About the MS-710 Scanner .................................................. vii About This Manual .............................................................. vii Keystroke Entries ...............................................
Chapter 4 Operational Commands Read Rate ......................................................................... 4-3 Trigger.............................................................................. 4-5 Scanner ............................................................................ 4-6 Counters ........................................................................... 4-7 Master Label ...................................................................... 4-9 Reset .............................
List of Illustrations Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure Figure 1-1 Typical Setup (without IB-105) .............................................. 1-2 1-2 Auxiliary Port Setup .............................................................. 1-2 1-3 Daisy Chain Setup ............................................
List of Tables Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table Table vi 1-1 Read Ranges and Scan Widths .......................................... 1-3 1-2 RS-232 15-pin External Cable Connections ......................... 1-6 1-3 RS-422/485 External 15-pin Cable Connections ................... 1-6 2-1 Symbology Identifier Option Values ................................. 2-31 3-1 Summary of Serial Configuration Commands ....................
About the MS-710 Scanner The MS-710 is a programmable, fixed-mount, high-speed scanner that reads and decodes a wide variety of bar code labels using a 10-sided spinning mirror to project laser beams over a wide scan angle. About This Manual This manual provides complete information on setting up and installing the MS-710 scanner. Chapter 1 provides overall step-by-step instructions for setting up and installing the MS-710 scanner with specific “go to” references to other chapters and appendices.
Warning and Caution Summary For connection to a UL Listed direct plug-in power unit marked Class II and rated at 5 VDC at 2 amps or greater. European Models must use a similarly rated Class I or Class II power supply that is certified to comply with the standard for Safety EN 60950. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
Safety Labels These Class II labels are located on the MS-710 scanner. LASER LIGHT IS EMITTED FROM THIS APERTURE AVOID EXPOSURE M A N U FA C TU R E D S ER I A L N U M B ER VO R S I CH T CA U TI O N L A S E R L I G HT D O N O T S T A RE I NT O B E A M. L IS T ED UL L A S E R S T RA H L U NG , W E N N A B DE CK U NG G E Ö F F NE T . NI CHT I N DE N S T R A HL B L I CK E N 1 95 0 4 K6 8 FI S ® ® LASERSTRAHLUNG NI CHT IN 6 7 0 n m L A S E R D I O DE 1 .
Zulassungen Dieses Gerät wurde für die Zulassung durch die folgenden Organisationen und in Übereinstimmung mit deren Normen entwickelt: CDRH (Center for Devices & Radiological Health; Zentrum für Geräte und Strahlenschutz) UL (Underwriters Laboratories, Inc.
Achtung/Warnung: WARNUNG Die Verwendung von Steuerelementen, Anpassungen oder das Ausführen von Prozeduren, die von den hier beschriebenen abweichen, kann dazu führen, daß Sie gefährlichen Laserstrahlen ausgesetzt werden. WARNUNG Der Scanner MS-710 enthält keine Teile, die durch den Benutzer gewartet werden können. Durch Öffnen des Abtastkopfes wird die Gewährleistung von Microscan Systems nichtig. Außerdem können Sie einer Laserdiodenleistung von bis zu 5 mW ausgesetzt werden.
xii MS-710 Scanner User’s Manual
1 Quick Start Chapter Contents Step 1 - Plan Scanning System .............................................................. 1-2 Step 2 - Attach Cabling ........................................................................... 1-3 Step 3 - Consider Read Range and Scan Width .................................... 1-4 Step 4 - Position Scanner ....................................................................... 1-5 Step 5 - Mount Scanner ....................................................................
Step 1 Plan Scanning System Before installing the MS-710 scanner you should sketch out a diagram of your scanning system, showing equipment, connector and cable types, and cable lengths. (For wiring information, see “Attach Cabling” on page 1-3.
Attach Cabling The MS-710 has a 15-pin internal connector with a standard, hard-wired 3-foot cable terminating in an external 15-pin connector. Since +5 VDC power supply is included in the single cabling assembly, the scanner cable should not exceed 3 feet. RS-232 cabling from the 15-pin connector to the host can be up to 47 feet provided it does not include 5 VDC input.
1–Quick Start Chapter 1 Step 3 Consider Read Range and Scan Width a) Verify that the labels to be scanned will pass through the readable range. Optimum decode rates can be expected at the center of the depth of field (halfway between minimum and maximum read range) for a given bar code density. b) Note the approximate scan width at the specific distance the labels will be passing through. Read range is the distance from the scanner in which a passing label can be read.
Position Scanner a) Position the scanner in a place devoid of sunlight, bright lights, or laser light from other sources. b) Avoid excessive tilt,skew, or pitch. Maximum tilt is determined by label characteristics and the number of scans required. Maximum skew is ±40°; maximum pitch is ±50°.1 c) Pitch label slightly to avoid specular reflection, the return of direct, non-diffused light.
1–Quick Start Chapter 1 Step 5 Mount Scanner The MS-710 scanner can be top or bottom mounted depending on your model type. Use the measurements provided in figure 1-8 to locate centers of mounting holes and drill as specified: Top mount – drill two 0.16 inch (4.06 mm) diameter holes. Bottom mount – drill two 0.203 inch (5.16 mm) diameter holes. 0.53" (1.35 cm) 3.47" (8.81 cm) 2.50" (6.35 cm) 0.519" (1.32 cm) 1.206" (3.06 cm) Optional Mounting Holes (2 places) #6-32 x 0.29 deep Top cable 2.59" (6.
Establish Communications To set up the host communications, from the ESP™ main menu: a) Select Serial Port and Configure from the pull-down menus. b) Make the port selection from Serial Port options COM 1 through COM 8. (Usually COM 2 for desktops and COM 1 for laptops.) c) Click the Auto Baud button down menu. or select Auto Baud from Serial Port pull ESP will cycle through the various baud, etc.
1–Quick Start Chapter 1 Step 7 Configure Scanner You can make changes from the default settings by modifying individual ESP menus and saving those changes to the scanner. For explanations of menu configuration settings, see Chapter 2, “Menu Configuration.” Note: When you save communications changes to the scanner, serial port settings are automatically matched.
Test Decode Rate with Static Label a) Position the label in front of a functioning scanner, click on the Utilities button, and in the Read Rate dialog either click the Percent or Decodes/Sec radio button and click Start, or enter the or command from the host terminal to launch test. b) Observe rate. (The percent or decode rate is added to the front of the data string.) c) End the test by clicking the Stop button or the command.
1–Quick Start Chapter 1 Step 9 Test Decode Rate with Moving Label Once you’ve mounted the scanner and established a good read rate with a stationary label, you can repeat the test with a moving label. To ensure reliable scanning, you need to apply a minimum of five scans to each label. This means that all variables, label speed, read range, etc. are significant.
1–Quick Start Step 10 Increase the Number of Scans Where possible, you can increase the number of scans a label receives by: • Increasing label height • Decreasing label length • Increasing scan rate (spinning mirror speed) • Slowing label speed See Appendix F, “Number of Scans,” on page A-11.
1–Quick Start Chapter 1 1-12 MS-710 Scanner User’s Manual
Chapter 2 Menu Configuration 2–Menu Config. Chapter Contents About the Menu Configuration Program .................................................. 2-2 Communications Menu............................................................................ 2-5 Protocol Menu ....................................................................................... 2-12 Operation Menu..................................................................................... 2-16 Code Type Menu .............................
Chapter 2 About the Menu Configuration Program Pull down menus Toolbar 2–Menu Config. An “Untitled...” file loads with the original default settings Access configuration menu changes here. Emulation window displays echoed commands and data Also titled “Current Menu/Scanner Settings” following a save or upload You can also access the configuration menus by clicking any item here.
About the Menu Configuration Program Saving Settings to the Scanner You can access the SEND/RECV box by clicking it directly from the Current Settings window, or from within any of the configuration menus, click Apply. To save changes to the scanner, select Apply in the menu pages. (Clicking OK on these menu pages will only update the items in the Current Settings window.
Chapter 2 Loss of Communications Making changes to communications parameters such as Baud Rate, Parity, Stop Bits, LRC, etc. without corresponding changes in linked device(s) can result in the loss of menu access. If this should occur, try ESP Auto Baud command or restore defaults. Caution: Defaulting the scanner will reset all standard configuration parameters except Scanner Type, Scans per Second, and Gain Adjustment to their original default values. 2–Menu Config.
Communications Menu Communications Menu 2–Menu Config. Figure 2-2 Communications Menu Structure Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons. Note: When you make and save menu communications changes to the scanner, those changes will automatically be matched in the Serial Port pull-down options.
Chapter 2 Baud Rate Default: 9600 Options: 300, 600, 1200, 2400, 4800, 9600, 19.2K, and 38.4K Allows you to set the number of bits transmitted per second. 2–Menu Config. Parity Default: Even Options: Even, Odd, None Allows you to select an error detection routine in which one data bit in each character is set to 1 or 0 so that the total number of 1 bits in the data field is even or odd.
Communications Menu Transparent Used to batch data from the aux port to the host. The scanner buffers data from the aux port and displays the keyed data on the aux port. The scanner transmits aux port data to the host when a label is scanned or a carriage return is entered from the aux port. Data Initiated from the Aux Port Aux. Port Host 2–Menu Config. • Aux Port data is passed through to the host whenever a return key is pressed at the aux port or whenever bar code data is sent.
Chapter 2 Half Duplex Mode In half duplex mode all aux port data and bar code data is sent directly to the host. Bar code data is displayed on the aux port screen at the same time the data is sent to the host. Data initiated from the Aux Port 2–Menu Config. • Aux Port data to the host is ignored if the scanner is in a polled mode. • Aux Port data or scanned data is sent to the host whenever it is received. Aux. Port Host • Aux Port data is not echoed.
Communications Menu Full Duplex Mode In full duplex mode, all aux port data and bar code data is sent directly to the host. Bar code data is not displayed on the aux port screen. Data initiated from the Aux Port • Aux port data is passed directly through to the host whenever it is received by the scanner, unless the scanner is in a polled mode, in which case the data will be ignored. Aux. Port Host • Aux Port data is not echoed. 2–Menu Config.
Chapter 2 Daisy Chain Mode 2–Menu Config. In a daisy chain application, a primary scanner is linked directly to the host and succeeding scanners are linked in tandem. The source of transmitted data can be identified by enabling and assigning daisy chain IDs to each scanner. Host Succeeding Scanner(s) Just before sending a noread message, the primary scanPrimary Scanner ner checks its auxiliary port for data. It waits a maximum of 20 ms.
Communications Menu Daisy Chain ID Status Default: Disabled Options: Disabled, Enabled. Daisy Chain ID Default: 1/ Options: Any one or two ASCII characters. MS-710 Scanner User’s Manual 2-11 2–Menu Config. Each scanner enabled for Daisy Chain can be assigned a one or two character Daisy Chain ID that appears in front of the data decoded by that scanner. Enable/Disable and Length must be the same in all scanners.
Chapter 2 Protocol Menu 2–Menu Config. Protocols define the sequence and format in which information is transferred between devices. Figure 2-3 Protocol Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons.
Protocol Menu host, when ready, responds with a CTS (clear-to-send) and the data is transmitted. CTS and RTS signals are transmitted over two dedicated wires (pins 6 and 10) as defined in the RS-232 standard. Point-to-Point with XON/XOFF (Transmitter On/Off) Used only with RS-232. This selection enables the host to send a single byte XON or XOFF transmission command of start (^Q) or stop (^S). If an XOFF has been received from the host, data will not be sent to the host until the host sends an XON.
Chapter 2 2–Menu Config. From Host Default: Disabled Options: Disabled, Enabled Messages sent to the host from the MS-810 will always include the scanner’s defined protocol. The status of From Host determines if messages sent to the MS-810 from the host must also include the defined protocol. If From Host is disabled, the defined protocol is not included. If From Host is enabled, the defined protocol must be included. (See “Explanation of the From Host option,” on page 3-7 for further information.
Protocol Menu Preamble (enable/disable) Default: Disabled Options: Disabled, Enabled (within any protocol) Allows you to enable or disable the preamble character(s). Postamble Default: ^M^J Options: Any ASCII character, including control characters. Postamble (enable/disable) Default: Enabled Options: Enabled, Disabled (within any protocol) Allows you to enable or disable the Postamble character(s).
Chapter 2 2–Menu Config. Operation Menu Figure 2-4 Operation Menu Structure Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons.
Operation Menu label is in the read range of the scanner. When To Output options have no affect on Continuous Read. Continuous Read is useful in testing label or scanner functions. Continuous Read 1 Output Note: Enabling Continuous Read 1 Output when Number of Labels (page 2-21) is set to any number greater than one will cause Number of Labels to default back to one.
Chapter 2 External Edge As with Level, Edge allows a read cycle to be initiated by a trigger signal from an object detector when it detects the appearance of an object (rising edge). But unlike Level mode, the removal of an object (falling edge) does not end the read cycle. With Edge enabled, the read cycle ends with a good read output, a timeout, or a new trigger (figure 2-6 on page 2-18). 1 2–Menu Config.
Operation Menu Timeout Can end the read cycle after a specified period of time, and if no label has been read, causes a noread message, if enabled, to be transmitted. With either External Edge, Serial Data, or Serial Data & Edge enabled, a timeout ends the read cycle. With External Level enabled, the read cycle does not end until the falling edge trigger occurs, and the next read cycle does not begin until the next rising edge trigger.
Chapter 2 Serial Trigger Character Default: ^]. Corresponds to displayed in the menu. Options: Any single ASCII character, including control characters, except NUL (00H), an existing host command character,1 or an on-line protocol character. 2–Menu Config. Note: Serial Data (page 2-18) or Serial Data & Edge (page 2-18) must be enabled for Serial Trigger Character to take effect. “N/A” is displayed in the menu when all other triggering modes are enabled.
Operation Menu Good Decode Reads Default: 1 Options: 1 to 31 Allows you to select the number of good reads (from 1 to 31) required per label before a good decode output. Note: Be sure to set the value within the determined scan rate for the scanning setup so that the scanner is capable of scanning a label the required number of times.
Chapter 2 Multilabel Separator Default: , (comma) Options: Any available ASCII character, except NUL. Allows you to choose the separator character to be inserted between each label. Note: UPC separator character, if used, must be different than the multilabel separator. 2–Menu Config.
Operation Menu Sequential Instructs the scanner to compare labels or portions of labels for sequential numbers in ascending or descending order. There are four ways to load a master (see “Master Label Entry” on page 2-24). In every case sequencing occurs after each match. See “Match Code Type” on page 3-11 for special noread and mismatch sequences. Wild Card Allows you to enter user defined wild card characters in the master label.
Chapter 2 Match Length Default: 1 Options: 1 to 31 2–Menu Config. Defines the length of the character string that will be compared with that of the master label when Match Code Type is set to Enabled or Sequential and Match Start position is set to 1 or greater. For example, If Match Length is set to 6 in a 10 character label, and Match Start Position is set for 2, only the 2nd through 7th characters (from left to right) will be compared.
Code Type Menu Code Type Menu 2–Menu Config. Figure 2-8 Code Types Menu Structure Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons. Code 39 Default: Enabled Options: Enabled, Disabled Check Digit Default: Disabled Options: Disabled, Enabled Code 39 is self-checking and does not normally require a check digit.
Chapter 2 Check Digit Output Default: Disabled Options: Disabled, Enabled When enabled, the check digit character is sent along with the label data. When disabled, label data is sent without the check digit. 2–Menu Config. Large Intercharacter Gap Default: Disabled Options: Disabled, Enabled Allows the scanner to read labels with gaps between bar code characters exceeding three times the narrow element width.
Code Type Menu Start & Stop Output Default: Enabled Options: Enabled, Disabled Allows the start and stop characters to be transmitted with bar code data. Large Intercharacter Gap Default: Disabled Options: Disabled, Enabled Fixed Code Length Default: Disabled Options: Disabled, Enabled Used to increase data integrity by ensuring that only label length will be accepted. Code Length Default: 10 Options: 1 to 64 Note: Fixed Code Length must be enabled for Code Length to take effect.
Chapter 2 Interleaved 2 of 5 Default: Disabled Options: Disabled, Enabled 2–Menu Config. Because I 2 of 5 is a continuous code, it is prone to substitution errors. Hence, a code length must be defined and a bar code label containing an even number of digits must be used. It is also recommended that a Modulus 10 check digit be used to ensure the best possible data integrity. Check Digit Default: Disabled Options: Disabled, Enabled I 2 of 5 uses a Modulus 10 check digit.
Code Type Menu UPC Default: Disabled Options: Disabled, Enabled When enabled, the scanner will read UPC version A and UPC version E only. EAN Disabled Options: Disabled, Enabled Note: UPC must be enabled for EAN to take effect. When EAN is enabled, the scanner will read UPC version A, UPC version E, EAN 13, and EAN 8. It will also append a leading zero to UPC version A label information and transmit 13 digits. If you do not want to transmit 13 digits when reading UPC version A labels, disable EAN.
Chapter 2 Note: If additional labels—other than the main or supplemental—will be read in the same read cycle, Number of Labels should be set accordingly. Separator Default: Disabled Options: Disabled, Enabled 2–Menu Config. Allows you to insert a character between the standard UPC or EAN code and the supplemental code when Supplementals is set to Required. Separator (character) Default: , (comma) Options: Any ASCII character.
Code Type Menu Symbology ID Default: Disabled Options: Disabled, Enabled When enabled, the scanner reads and outputs a three character prefix: 1. ] (close bracket character) indicating the presence of a symbology ID 2.
Chapter 2 2–Menu Config. Output Menu Figure 2-9 Scanner Output Menu Structure Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons. Note: Bar code output data and noread messages are transmitted on pin 2 under conditions set principally in Bar Code Output.
Output Menu Output On Default: Mismatch or Noread Options: Mismatch or Noread, Match, Mismatch, Noread Note: If Output On is set to Mismatch or Noread, Match, or Mismatch, a relay pulse will not output unless Match Code Type (page 2-22) is not disabled and a master label is downloaded into memory (page 2-21). Note: To send a pulse on a Good Read, enable Match and disable Match Code Type (see “Match Code Type” on page 2-22).
Chapter 2 # Before Output Default: 1 Options: 1 to 255 Allows you to determine the number of consecutive hits (matches, mismatches, or noreads as configured by Output On) that must occur before a relay pulse will be sent from Relay-1 or Relay-2. 2–Menu Config. For example, if # Before Output is set to 3 and Relay-1 is set to Noread, then relay-1 will not fire until 3 consecutive noreads have occurred. Note: The hits must be consecutive and each one must match the configured Output On condition.
Output Menu End of Read Cycle Causes bar code data output to be delayed until the end of the read cycle. Noread Message Default: NOREAD Options: Up to seven ASCII characters (except NUL) Noread Output (enable/disable) Default: Enabled Options: Enabled, Disabled Note: If Noread Output is enabled, the noread message will only output if Bar Code Output (page 2-34) is also enabled and MS-sensor Status (page 2-40) is disabled. Allows you to enable or disable the noread message.
Chapter 2 K Command Options 2–Menu Config. Echo K Command Default: Disabled Options: Disabled, Enabled When enabled, serial configuration commands (“K” commands) will be processed and the new string for that command will be echoed back to the host. If an invalid command is sent to the host, the scanner will echo back the current setting of that command. For example, if the current Noread Message is “NOREAD” and is entered, the scanner will echo back: .
Output Menu Partial Output Default: Disabled Options: Disabled, Enabled When enabled, allows you to pre-select portions of labels to be transmitted by the scanner. In multilabel mode, partial output is performed on each separate label. For example, if Start Position is set to 3 and Partial Length is set to 5, the following labels are transmitted as follows: 2–Menu Config.
Chapter 2 Setup Menu 2–Menu Config. The Setup menu allows you to set specific bar code scanning parameters that affect decode rate, and if equipped with the MS-sensor, allows you to determine why the scanner did not decode a specific label within the triggered read cycle. Figure 2-10 Setup Menu Structure Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons.
Setup Menu Factory Settings Scans per Second Default: 50. Corresponds to 500 (scans per second) displayed in the menu. Options: 30 to 55. Multiply the number to be entered by 10 for actual scans per second. Allows you to set the number of scans per second by controlling mirror motor speed. The range is 300 to 550 scans per second. Default: 45 (nominal) Options: 10 to 255 2–Menu Config.
Chapter 2 MS-sensor Parameters1 Default: Disabled Options: Disabled, Enabled 2–Menu Config. Note: See “MS-Sensor” on page A-24. The specular sensor—once you adjust it to your environment—determines why the scanner fails to decode a label within the triggered read cycle. It then outputs the appropriate message to you. The following three conditions are possible: 1. the object’s bar code is defective (you receive a bad bar code message) 2.
Setup Menu No Barcode Status Default: Disabled Options: Disabled, Enabled Allows you enable or disable the no label message. No Barcode Message Default: NOLABEL Options: Up to seven ASCII characters (except NUL) Bad Barcode Status Default: Disabled Options: Disabled, Enabled Allows you to enable or disable the bad label message. Bad Barcode Message Default: BADCODE Options: Up to seven ASCII characters (except NUL) Allows you to define the message that is sent when a label is defective.
2–Menu Config.
Chapter 3 Serial Configuration Chapter Contents Most of the configuration changes to the scanner that can be made in the menu can also be accomplished by command strings from the host. Serial command strings are entered from an ASCII terminal. As with menu configuration commands, serial configuration commands relate to the initial scanner setup. To quickly reference the serial configuration commands, see table 3-1 on page 3-2. MS-710 Scanner User’s Manual 3-1 3–Serial Config.
Chapter 3 Table 3-1 Summary of Serial Configuration Commands Function Comm. Port Protocol 3–Serial Config.
About Serial Configuration About Serial Configuration Serial Configuration Command Format The format for a serial configuration command is, Where: • Less than < and greater than > symbols are included as part of the commands. • “parameter,” as used here, are those character(s) that precede the data. • The “initializing command” or is sent after configuration is complete. resets and saves for power up. initializes the change to RAM.
Chapter 3 Initializing serial configuration commands To ensure that a serial configuration command will take effect, you need to follow it with one of the operational commands below: To reset but not save changes for power-on To reset and save changes for power-on For example, to change Baud Rate and reset without saving changes for power-up, enter . To change Baud Rate and reset, saving the changes to NOVRAM, enter .
Communications Port Commands Communications Port Commands Host Port Note: Changes made in the scanner’s communications parameters such as baud rate, parity, stop bits, LRC, etc., must be matched in the other device(s) or communications will be lost. If this occurs, default the scanner with the Restore/Save Default Configuration for Power-on command as described on page 4-12.
Chapter 3 Protocol Commands If selecting one of the options from 0 to 4 (Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, or Polling Mode D), use this format: Format: 1 protocol: 0 = Point-to-Point 1 = Point-to-Point with RTS/CTS 2 = Point-to-Point with XON/XOFF 3 = Point-to-Point with RTS/CTS and XON/XOFF 4 5 6 7 = = = = Polling Mode D Multidrop (requires address) User Defined User Defined Multidrop Example: To change the
Protocol Commands Note: Address, can be assigned any ASCII character except a null. Control characters are used to define RES through NAK (except Address). Table 3-2 lists the control characters used for these data fields. (Refer to Table A-4, “ASCII Table with Control Characters,” on page A-7 for more information.
Chapter 3 Table 3-2 Protocol Commands Protocol Command (Mnemonic displayed on Microscan menu) Control Characters (Entered in menu or serial command) Effect of Command RES REQ EOT STX ETX ACK NAK ^D ^E ^D ^B ^C ^F ^U Reset Request Reset Start of Text End of Text Acknowledge Negative Acknowledge Preamble 3–Serial Config. Format: status: preamble character(s): 0 = Disabled Enter one or two preamble characters except a null (00H). Default is ^M.
Protocol Commands Response Timeout Format: 1 timeout setting: Default is 12 (ms). Any number from 0 to 65,000 (a zero causes an indefinite wait). Example: To change Response Timeout to 30 ms, enter . Intercharacter Delay Format: time interval (between characters in milliseconds): Default is 0. Any number from 0 to 255. Example: To change Inter-character Delay to 30 ms, enter . 3–Serial Config. 1. For use with Microscan’s IB-105.
Chapter 3 Operation Commands Triggering Mode Format: trigger filter time (in 128 µs increments): triggering mode: Default is 78 (9984 µs; approx. 10 ms) 0 = Continuous Read Any number between 2 and 255. 1 = Continuous Read 1 Output 2 = External Level 3 = External Edge 4 = Serial Data 5 = Serial Data & Edge Example: To select External Edge, enter . 3–Serial Config.
Operation Commands External Trigger Level Format: external trigger level (initiates a read cycle): 0 = Negative 1 = Positive Example: To change External Trigger Level to Negative, enter . Good Decode Reads Format: number of reads (before a good decode): Default is 1. Any number from 1 to 31. Example: To change Good Decode Reads to 3, enter .
Chapter 3 Example: To enable Match Code Type for Sequential, set Sequential Matching for Decrement, and enable Sequence on Every Mismatch, enter . New Master Pin Format: status: 0 = Disabled 1 = Enabled Example: To enable New Master Pin, enter . Number of Labels Format: 3–Serial Config. number of labels: Default is 1. Any number from 1 to 6 multilabel separator: Default is ,.
Code Type Commands Code Type Commands Code 39 Format: status: 0 = Disabled 1 = Enabled check digit status: 0 = Disabled 1 = Enabled large inter-character gap:fixed code length status: 0 = Disabled 0 = Disabled 1 = Enabled 1 = Enabled check digit output status: 0 = Disabled 1 = Enabled code length: Default is 10. Any number from 1 to 64.
Chapter 3 I 2 of 5 Format: status: 0 = Disabled 1 = Enabled check digit status: 0 = Disabled 1 = Enabled check digit output status: 0 = Disabled 1 = Enabled code length #2: code length #1: Default is 6. Default is 10. Zero or any even Zero or any even number from 2 to 64. number from 2 to 64. Example: To set Fixed Code Length #1 to 8 and Fixed Code Length #2 to 4, enter or . 3–Serial Config.
Code Type Commands Narrow Margins/Symbology ID Format: narrow margins status: 0 = Disabled 1 = Enabled symbology ID status: 0 = Disabled 1 = Enabled Example: To enable Narrow Margins, send . Autodiscriminate Format: Enables all available symbology types and disables OMR. You may individually disable/ enable each symbology type also. Note: For maximum scanning speed, enable only those symbologies used in the application.
Chapter 3 Scanner Output Commands Relay-1 Format: output on (mode): 0 = Mismatch or Noread 1 = Match (or Good Read) 2 = Mismatch 3 = Noread polarity: pulse width: 0 = Negative Default is 5 (50 ms). 1 = Positive 0 to 255 # before output: Default is 1. 1 to 255 Example: To set Relay-1 to Mismatch and change Pulse Width to 40 ms, and # Before Output to 5, enter or . Relay-2 3–Serial Config.
Scanner Output Commands Noread Message Format: status: noread output: 0 = Disabled Default is NOREAD. 1 = Enabled Any ASCII string up to 7 digits. Example: To enable Noread Message and send the message “FAIL,” enter .
Chapter 3 Serial Verification Format: serial command echo status: 0 = Disabled 1 = Enabled serial command beep status: 0 = Disabled 1 = Enabled control/hex output: 0 = Control 1 = Hex Example: To enable Serial Command Echo Status and Beep Status, and Hex output, enter Scanner Output Status Request Format: 3–Serial Config. Returns status of each command in the Scanner Output group.
Scanner Setup Commands Scanner Setup Commands Scanner Type Important: Entering the command (see “Save Scanner Type, SPS, and Gain for Power-on” on page 4-12) will save the current setting for Scanner Type to NOVRAM. Format: scanner density type:* 0 = Standard 1 = High Density * The “default” will depend on the type of scanner. Example: Do not change the scanner density type unless instructed to do so by a Microscan representative. Scans per Second 3–Serial Config.
Chapter 3 MS-sensor Parameters 1 Format: MS-sensor status: 0 = Disabled 1 = Enabled number of transitions: Default is 14. Any number from 3 to 255 reflectance threshold: Default is 40. Any number from 0 to 255. Example: To enable the sensor and set Number of Transitions to 20 and the reflectance threshold to 45, enter . Reflectance Setup Test 3–Serial Config.
Scanner Setup Commands No Barcode Message Format: no barcode status: 0 = Disabled 1 = Enabled message: Any ASCII string up to 7 characters. Default is NOLABEL. Example: To change the message to “NOCODE,” enter No Object Message Format: no object output status: 0 = Disabled 1 = Enabled message: Default is MISSING. Any ASCII string up to 7 characters. Example: To change the message to “VACANT,” enter . 3–Serial Config.
3–Serial Config.
Chapter Utilities 4 Chapter Contents Read Rate ............................................................................................... 4-3 Trigger ..................................................................................................... 4-5 Scanner ................................................................................................... 4-6 Counters .................................................................................................. 4-7 Master Label............
Chapter 4 Table 4-1 Summary of Operational Commands Type Read Rate Trigger Scanner Counters Master Label 4–Utilities Reset Checksum Code Types (not in ESP) 4-2 Cmd Result Decode Rate Test for Single Label Percentage Rate Test for Single Label Multilabel Decode Rate Test Multilabel Percentage Rate Test Extended Decode Rate Test for Single or Multilabel Exit any Rate Test Serial Trigger Character Enable Laser Scanning (Laser On) Disable Laser Scanning (La
Read Rate Read Rate Decode (and percentage) rates can vary dramatically due to the angle and location of the label in relation to the scan beam (or scan line, if using a scan head with a moving beam). These tests are very useful in aligning and positioning the scanning device during installation. Single Label Percent Test Clicking Single Label Percent and Start instructs the scanner to output the percent of decodes and any scanned label data.
Chapter 4 End Read Rate Test Clicking Stop ends both the Percent test and the Decodes/Second test for both single and multi-label. Associated serial command: Extended Mode Choosing any of the Read Rate modes along with Extended Mode affects both single and multiple labels. It instructs the scanner to output the transition data along with the data from the latest test command sent.
Trigger Trigger Clicking the Setup button will bring up the current trigger character. Next click Send to send the trigger character to the scanner and initiate a read cycle. The results, a good read or noread message, will be displayed on the blank screen below the button. Note: Your scanner must be in a point-to-point protocol for Trigger Setup to work. Note: You can define the trigger character in the Operations menu.
Chapter 4 Scanner Enable Laser Scanning Clicking the Laser On button or sending serial command will turn the laser on continuously. Note: Enable Laser Scanning does not relate to Laser On/Off command. Associated serial command: Disable Laser Scanning Clicking Laser Off will turn the laser off. This feature is useful during extended periods of time when no bar code labels are being scanned or the scanner is being configured.
Counters Counters Counter commands can be a numeric value from 00000 to 65,535. After reaching the maximum numeric limit of 65,535, an error message will be displayed and the counter will automatically rollover and start counting again at 00000. To obtain the cumulative total of counts after the rollover has occurred, add 65,536 per each rollover (the scanner does not keep track of the number of rollovers) to the current count.
Chapter 4 or, if Master Label is not enabled, the number of good reads since the last reset. This counter is always enabled, but will only work as a match count when Master Label is enabled. If Master Label is not enabled, this counter records the number of good reads. This count can be requested at any time. Associated serial command: Match Counter Reset Clicking Reset sets the Match Counter to 00000.
Master Label Master Label Enable Match Code Type Identical to “Match Code Type,” on page 2-22. Instructs the scanner to compare bar code labels being scanned with a master label and may under certain conditions send out a relay driver signal and update internal counters. If no master label has been entered, every decoded label will be a “mismatch” and will increment the mismatch counter by one. Enable Match Code Type is intended for use when the scanner is in a triggered mode.
Chapter 4 Request Master Label Information Immediately sends the master label information to the host. To prevent conflicts with outputting label data, first send the command (Disable Laser Scanning (Laser Off)). Associated serial commands: <)> Delete Master Label Information Deletes master label information that has previously been loaded by either <)XXXX)> Download Master Label Information Command or Store Next Label as Master Label command.
Reset Reset Important: All and commands must be followed by a one second delay before any operations, commands, or reads can be processed. Reset Initializes all serial configuration commands in RAM and resets all counters and operating parameters. Initialize RAM Note: Software Reset will reset all counters. You should record all data that you wish to save prior to sending this command.
Chapter 4 Reset/Save Parameters for Power-on Allows you to save all configuration command settings to NOVRAM. Numeric counter values and master label data are not saved by this command.1 Associated serial command: Restore/Save Default Configuration for Power-on Allows you to restore and save default settings for all but Scanner Type, Scans per Second, and Gain Adjustment to NOVRAM. RAM ROM NOVRAM Note: Power must be available to the scanner during the default procedure.
Checksum Checksum Part Number Displays software part number. Associated serial command: <#> Checksum Displays a four-digit hex number (corresponding to a given firmware version) used to verify a scanner’s EPROM. Associated serial command: Show Scanner Status Used to display certain operating parameters (NOVRAM, command, and operation status). After sending this command, the scanner immediately responds with , where “X” is an ASCII character between @ and ∆ on the ASCII table.
Chapter 4 3.
Code Type Commands Code Type Commands (Not in the Utilities Menus.) Autodiscriminate All Codes Enables the scanner to decode all available bar code types without changing scanner configuration settings. Note: For maximum scanning speed, enable only those bar code symbologies used in the application. Associated serial command:
Code 39 Only Allows only Code 39 labels to be read. Associated serial command: Codabar Only Allows only Codabar labels to be read.
4–Utilities Chapter 4 4-16 MS-710 Scanner User’s Manual
Appendices A Contents Appendix A — Scanner Specifications ..................................................A-2 Appendix B — Electrical Specifications .................................................A-5 Appendix C — ASCII Table ...................................................................A-7 Appendix D — Orientation .....................................................................A-8 Appendix E — Defaulting the Scanner ................................................
Appendix A — Scanner Specifications Mechanical -- MS-710 Light Source Length: 3.47” (88.1 mm) Height: 1.32” (33.5 mm) Type: Laser Diode Output Wavelength: 670 nm nominal Safety Class: CDRH Class II Operating Life: 50,000 hours @25°C (10,000 hours @50°C) Laser Cooling: Thermostatically controlled thermocooler Width: 2.59” (65.8 mm) Weight: 9 oz. (255 g) 0.53" (1.35 cm) 3.47" (8.81 cm ) 2.50" (6.35 cm ) Scanning Parameters 0.519" (1.32 cm ) 1.206" (3.
Scanner Specifications Status Lights Red 5V power is present Ready Green Scanner is ready to read Good Read Green Label is qualified as good In menu status both green lights will flash on and off. Optical Characteristics Optics: Coaxial Operating range (depending on label type and size): High Density: 2 to 4 inches Low Density: 2 to 10 inches Right-angle Down: 1 to 8.5 inches Scan width: 7 inches maximum Skew: ±40° maximum Pitch: ±50° maximum Label contrast: 25% min.
FIS Options Each MS-710 model is identified by an FIS (final instruction sheet) number that refer to a set of options.
Electrical Specifications Appendix B — Electrical Specifications Power Supply Requirements: Input +5 VDC ±4% regulated @ 480 mA with 200 mV p-p max. ripple. Table A-2 applies to those FIS options that use RS-232 communications. Table A-3 applies to those FIS options that use RS-422/485 options. The affected pins are 2, 3, 6, and 10.
Ground and Shield Considerations Proper grounding is necessary for operator safety, noise reduction, and the protection of equipment from voltage transients. Buildings, including any steelwork, all circuits, and all junction boxes must be grounded directly to an earth ground in compliance with local and national electrical codes. RS-232 signals have a common signal ground (pin 4). Pin 4 is normally connected to pin 13 (chassis ground) in the scanner; however, under certain conditions (e.g.
ASCII Table Appendix C — ASCII Table Table A-4 ASCII Table with Control Characters Hex Mne Ctrl Dec Hex Ch Dec Hex Ch Dec Hex Ch 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 00 01 02 03 04 05 06 07 08 09 0A 0B 0C 0D 0E 0F 10 11 12 13 14 15 16 17 18 19 1A 1B 1C 1D 1E 1F NUL SOH STX ETX EOT ENQ ACK BEL BS HT LF VT FF CR SO SI DLE DC1 DC2 DC3 DC4 NAK SYN ETB CAN EM SUB ESC FS GS RS US ^@ ^A ^B ^C ^D ^E ^F ^G ^H ^I ^J ^K ^L ^M ^N ^O ^P ^Q ^R ^S
Appendix D — Orientation If the bar code label bars are parallel to the direction of travel, as shown in figure A-5, the label is said to be in a ladder orientation; if the bars are perpendicular with the direction of travel, the label is said to be in a picket fence orientation (figure A-6). Ladder Orientation In general, depending on label size and speed, ladder orientation is preferable because different portions of the label are scanned as the label goes by.
Orientation Angled Picket Fence Orientation The problems associated with picket fence can be minimized by slightly tilting the scan line (figure A-7). This allows a larger portion of the label to be scanned and increases the label placement area by “simulating” the ladder orientation. This of course is impractical with a short bar code height.
Appendix E — Defaulting the Scanner Defaulting the scanner resets all scanner configuration parameters except Scanner Type, SPS, and Gain Adjustment to their original default values. These values are saved to NOVRAM for power-on by entering the command (Save Scanner Type, SPS, and Gain for Power-on). RAM NOVRAM ROM Defaulting is necessary if: • You wish to quickly restore default settings to the configuration program after making some temporary changes.
Number of Scans Appendix F — Number of Scans To ensure reliable scanning, apply a minimum of five scans to each label. Use the formulas below to calculate the number of scans that your label will receive. For definitions of terms, see inset on page A-12. If the number of scans you derive from one of these calculations is less than the minimum for your application, plug in the minimum number of scans (5) and solve for another parameter that might be changed, such as label speed or scans per second.
Angled Picket Fence Calculation Direction of label travel The number of complete scans for angled picket fence is calculated the same as that for picket fence, with the exception that the scan width is shortened in proportion to scan tilt. Scan Line Figure A-11 Angled Picket Fence Orientation Definitions Scan Rate (SR) is the number of scans per second that a given scanner is capable of emitting.
Number of Scans After making changes, repeat “Number of Scans” on page A-11 and step 9, “Test Decode Rate with Moving Label,” on page 1-10. Range Adjusting the label’s range, if possible, is one of the quickest and most effective ways to improve decode rates. However, in some applications you may need to select a less than optimum range, or one that is beyond the fringes of the ranges listed in table A-1 on page A-3.
Appendix G — Troubleshooting Operational Tips Do: • Check inputs (label speed, length, height, etc.) to ensure the desired number of scans per label. • Avoid excessive tilt, pitch, and skew of the bar code label. • Check the label for readability by doing a decode rate test. If there is any question about the label's readability, contact your Microscan representative. • After changing any parameter that might affect decode rate, repeat decode rate test.
Troubleshooting Table A-5 Troubleshooting Problem Getting an “Upload from Scanner has timed out!” message when clicking the Receive from Scanner command. No data or decode rate in decode rate test () Possible Cause Host cable defective or not wired properly Scanner configuration settings do not match host's Check cable connections and wiring. (See “Attach Cabling” on page 1-3.) Reset scanner to default and/or match host settings with scanner's.page A-10.
Table A-5 Troubleshooting Problem Possible Cause Decode rate okay but not getting any output (including noreads) in triggered mode Object detector not positioned properly End of Read Cycle not properly defined Scanner not triggered when in external mode Decode rate okay but getting only noread messages in triggered mode Previous label read, but subsequent label will not read Scanner not entering read cycle in triggered mode (trigger not working) Ensure read cycle is active during the time the label i
Troubleshooting Table A-5 Troubleshooting Problem Possible Cause MS-sensor display- Reflectance Threshold or ing incorrect error Minimum Label Transitions messages set incorrectly “Bad bar code” message Poor bar code quality or incorrect label position. “No bar code” mes- No label is present. sage “No object” No object is present. message Solution See “MS-sensor Parameters” on page 240.
Appendix H — Using the IB-105 P/N: 99-420001 The IB-105 is a separately sold item that provides additional functionality to the MS-710. Specifically: • Default switch • A normally open relay • RTS/CTS • RS-485 Multidrop • New master switch • RS-232/RS-422 Mode switch • Aux port • Power switch The IB-105 is available in five options (see Table A-11, “IB-105 Options,” on page A20). Mechanical Environmental Length: 4.5 in. (114.3 mm) Width: 3 in. (76.2 mm) Height: 1.5 in. (38.
Using the IB-105 Table A-9 25-pin D-sub Host Connector Table A-6 6-pin Trigger Connector Pin a Function Pin Trigger in to pin 9 of scanner connector 1 2 Relay contact normally open (up to 1watt) 2 Host RS-232 TXD (RS-232) 3 VCC (+5V) 3 Host RS-232 RXD (RS-232) 4 VBB (+12V, if using Microscan’s standard object detectora) 4 5 Ground Host RS-232 RTS or under software control aux RS-232 TXD, or RS-485 transmitter enable (out) 6 Relay common 5 Host RS-232 CTS or under software control aux
Table A-11 IB-105 Options P/N 99-420001-XX Description –01 RS-232 with RTS/CTS or RS-485 –02 RS-232 with RTS/CTS or RS-422 –03 RS-422 with Aux RS-232 or RS-232 –04 RS-422 (MS-710) to RS-232 –05 RS-422 LAN to RS-232 converter Internal Dip Switches 8 7 6 5 4 = ON 3 2 1 = OFF The IB-105 routes signals between the scanner and other devices.
Using the IB-105 Mode Switch Used in options -01 and -02 to toggle the host port between RS -232 and RS-485 or between RS-232 and RS-422, respectively. Options -03 and -04 require that the Mode switch be pushed IN for proper operation. 25-pin (host) Connector The 25-pin (host) connector is similar to the MS-3000 host connector; therefore, applications that were set up for the MS-3000 will work with the IB-105.
OPTION -02: RS-232/422 with RTS/CTS or Second RS-232 Serial Port Use the Mode switch to toggle the host port between RS-232 and RS-422. When the Mode switch is OUT, the RS-232 signals go through to the 25-pin (host) connector. Note: The second RS-232 TXD pin is available in this mode. Scanner RS-232 IB-105 RS-232 Host • When the Mode switch is IN, the IB-105 takes RS-232 signals (TXD, RXD) and converts them to RS-422 levels.
Using the IB-105 OPTION -05: RS-422 LAN to RS-232 Terminal 1 In this configuration the Mode switch must be IN for proper operation. This is a special configuration to convert RS-422 signals back to RS-232. The main purpose for this configuration is to use two IB-105’s, one to convert the scanner communications to RS-422 and one to convert these signals back to RS-232 (this configuration) for long distance communications (up to 4000 feet).
Appendix I — MS-Sensor Note: The MS-sensor is an optional FIS feature. The MS-sensor is a patented specular sensor that interprets reflected light from the laser beam. Once adjusted to your environment, it indicates why the scanner might fail to decode a label within the triggered read cycle and outputs the appropriate message to you.
MS-Sensor Scan Width cm 20 in. 15 10 5 5 0 0 MS-sensor's specular detection area 5 10 15 20 5 Scan Beam 25 10 20 Specular Objects Diffused Objects 15 5 10 5 MS-710 Scan angle equals 60 degrees 0 0 cm in. Range Figure A-15 MS-sensor Range b) Determine the setting for Reflectance Threshold (the minimum reflective level at which an object is detected) by selecting a value that is one or two numbers greater than the value posted next to REF when nothing is in the scanner’s beam.
Start Trigger Event Scans until End of Read Cycle Does the scanner decode the label? Yes No Is MS-sensor Status enabled? No Scanner outputs the noread message Yes Scanner outputs the bad label message (if enabled) Yes Scanner outputs the no label message (if enabled) Yes Are the number of transitions greater than the user defined value? No Is the reflectance greater than the user defined threshold? No Appendices Scanner outputs the no object message (if enabled) Figure A-16 MS-sensor Flowch
Bar Code Symbology Appendix J — Bar Code Symbology Before choosing a bar code symbol, consider:1 • the type of information to be scanned • the length of the messages to be encoded • how and where labels are to be applied • label printer capabilities • • • • • scan speed range and beam width the space available on the object to be identified host software limitations the speed of the conveyor The Standard MS-710 supports the following five bar code symbologies.
Appendix K — Interface Standards Interface Standards, established by the Electronic Industries Association (EIA), specify such things as the signaling voltage levels, maximum cable lengths, and number of drivers. With Microscan devices, selection of interface is made by pin assignment and, in the case of the host communications, by software switching between RS-232 and RS-422. Microscan devices use RS-232, RS-422, and RS-485 multidrop.
Multidrop Communications Appendix L — Multidrop Communications This appendix describes the rules for setting up a concentrator or controller to communicate with a scanner in standard Multidrop protocol. The diagram to the right shows a typical Multidrop network in which 1 to 50 scanners can communicate with a host via an intermediary device, a concentrator or a controller. Polling Sequence Host Concentrator Data that is transmitted to the host (bar code data, noread messages, counters, etc.
• If the scanner transmits data to the concentrator and the concentrator responds with an ACK or NAK, but the scanner doesn’t receive the concentrator’s response, the scanner will timeout and transmit a REQ to the concentrator and request another response. If after three retries (the number of times it transmits a REQ to the concentrator) the scanner receives no response, it ends the transmission with a RES (reset).
Multidrop Communications Table A-12 Multidrop Address Characters Multidrop Address 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 Poll Character ASCII HEX ^\ ^^ SP " $ & ( * , .
Appendix M — Glossary of Terms Autodiscriminate. The ability to decode several different bar code symbologies without changing configuration. AWG. Abbreviation for American Wire Gauge. This is a standard measuring system for wire sizes, listed from 0 (biggest) to 40 (smallest). A 26 AWG is equivalent to 0.0159 in. (0.405 mm) in diameter. Bar Code. Data that has been encoded into an array of parallel bars and spaces of varying widths. Bar Code Density. Number of characters per inch or other unit of measure.
Glossary of Terms Label Length. Regardless of orientation, the measurement taken across the label's bars from one end to the other, including the quiet zone. Label Speed. The rate in inches or centimeters per second at which a label moves through the scan beam. Label Transitions. The transition of bars and spaces on a label, used by the MS-sensor to detect the presence of a label on an object. Ladder Label Orientation. A bar code label in which the bars are parallel to the label’s direction of travel.
Port. Logical circuit for data entry and exit. (One or more ports may be included within a single connector.) Primary Scanner. First scanner in a daisy chain mode and linked directly to the host and in tandem to succeeding scanners. Protocol. The rules for communication between devices, providing a means to control the orderly flow of information between linked devices. Quiet Zones. Specified “clear” (nonprinted) areas immediately before and after the bar code symbol.
Glossary of Terms Successive Scanner. Linked to the primary or preceding scanner in a daisy chain mode via its aux port to relay bar code data to the host. See “Daisy Chain.” Symbology. A set of bar code symbols, such as Code 39 or Code 128, that have special rules to define the widths and positions of bars and spaces to represent specific numeric or alphanumeric information. Tilt. Label (or scanner) rotation around the centerline of the scan beam. Timeout.
Appendices A-36 MS-710 Scanner User’s Manual
Index Symbols A Achtung/Warnung xi AGC Sampling 2-39 Angled Picket Fence Orientation A-9 Approvals vii As Soon As Possible 3-16 ASCII table (with control characters) A-7 Auto Baud 1-7, 2-4 Auto Gain Control 3-19 Autodiscriminate 3-15 Autodiscriminate (definition) A-32 Autodiscriminate All Codes 4-15 Aux Port Parameters 2-6 B Bad Barcode Message 2-41, 3-20 Bad Barcode Output 2-41 Bar Code Output 2-34, 3-16 Baud Rate 2-6, 3-5 Baud Rates A-2 Beep on K 2-36 Beeper 2-35, 3-17 Beeper Volume 2-35, 3-17 Index #
Index Index E EAN 2-29, 3-14 Echo K Command 2-36 Edge 2-18 Electrical Specifications A-5 End of Read Cycle 2-18, 3-10, 3-16 End Read Rate Test 4-4 ESP (using) 2-2 Extended Decode Rate Test A-24 Extended Mode 4-4 External Edge 2-18, 3-10 External Level 2-17, 3-10 External Trigger Level 3-11 F Factory Settings 2-39 FCC x FIS (final instruction sheet) A-4 Fixed Code Length (Codabar) 2-27 Fixed Code Length (Code 128) 2-30 Fixed Code Length (Code 39) 2-26 From Host 2-14 Full Duplex Mode 2-9 G Gain A-13 Gain
N Narrow Margins 2-30, 3-15 New Master Pin 2-21, 3-12 New Trigger 2-19, 3-10 No Barcode Message 2-41 No Barcode Output 2-41, 3-21 No Object Message 2-41, 3-21 No Object Output 2-41 Noise Interference A-6, A-28 Noread Counter 4-7 Noread Message 2-35 Noread Output 2-35, 3-17 Number of Labels 2-21, 3-12 Number of Scans 1-11, A-11 Number of scans calculation A-11 Number of Transistions 3-20 O Operating Environment A-2 Operating Life A-2 Operation Commands 3-10 Operation Menu 2-16 Operations Status Request 3-12
Index Index Reset/Read NOVRAM 4-11 Reset/Restore 4-11 Reset/Save Parameters for Power-on 4-12 Response Timeout 2-15, 3-9 Restoring Default Settings 2-3 Right-angle Down Scan Range/Width Graph A-3 RS-232 A-28 RS-232 15-pin External Cable Connections A-5 RS-422 A-28 RS-422/485 External 15-pin Cable Connections A-5 RS-485 A-28 S Safety Class A-2 Safety Labels ix Save Scanner Type, SPS, and Gain for Power-on 4-12 Saving Settings 2-3 Scan Range/Width Graph A-3 Scan Rate A-2, A-12 Scan Rate definition A-12 Scan
U Index UL x UPC/EAN 2-29, 3-14, A-27 User Defined 2-13, 3-6 User Defined Multidrop 2-14, 3-6 V Vorsichtsmaßnahmen x W Warning Summary viii Warnung xi Warranty Information ii Wavelength A-2 When to Output 2-34, 3-16 Wild Card 2-23 Wild Card Character 2-24 MS-710 Scanner User’s Manual I-5