User manual
11-22
EZ Match User Manual
Tube/Cap Mode
Tube/Cap Mode
Tube/Cap Direction
Right
The tube or cap points to the right.
Left
The tube or cap points to the left.
Tube/Cap Rotation
Static
The tube or cap is static while being read (i.e., not rotating). This is the
Decapper Gate
default.
Rotating
The tube or cap rotates while being read. This is the Divert Gate default.
Cap Unknown Qualify
As Soon as Possible (ASAP)
When this mode is selected, a “Cap Unknown” condition is qualified as soon as it is
detected and meets the qualification requirements.
End of Operation
When this option is selected and a “Cap Unknown” condition is detected, it will not qualify
until the end of the operation. This allows more opportunity for a “Cap Present” condition
to be detected.
Serial Cmd:
<
K533,tube/cap direction
,tube/cap rotation,cap unknown qualify,tube/cap
presence before output,cap unknown presence before output>
Default: Right
Options: 0 = Right 1 = Left
Serial Cmd: <K533,tube/cap direction,tube/cap rotation,cap unknown qualify,tube/cap
presence before output,cap unknown presence before output>
Default: Static (Decapper Gate)
Default: Rotating (Divert Gate)
Options: 0 = Static 1 = Rotating
Serial Cmd:
<
K533,
tube/cap direction,tube/cap rotation,
cap unknown qualify
,tube/cap
presence before output,cap unknown presence before output>
Default: End of Operation
Options: 0 = ASAP 1 = End of Operation