User's Manual
32
Network cable is in poor contact.
Reconnect the network cable or
replace with another one.
Severe environment interference.
Move the device away from the metal
shield of the environment.
Internal parts are damaged.
Return to MicronView or the local
distributor for service.
8
Blower does not work
when sampling.
System malfunction.
Shut down and restart the device.
Internal parts are damaged.
Return to MicronView or the local
distributor for service.
9
Error alarm or counting
alarm is without voice
prompt.
Audible alarm is not turned on.
Turn on the audible alarm in the
setting interface.
System malfunction.
Shut down and restart the device.
Internal parts are damaged.
Return to MicronView or the local
distributor for service.
Hardware
10
Device does not turn on.
Internal short circuit.
Observe the indicator light when
supply with AC adapter. If the light is
off, return to MicronView for service.
If the light is on, an internal short
circuit is not the issue.
The AC adapter does not work.
If the problem persists after plugging
in the AC adapter again, replace the
AC adapter.
Internal parts failure.
Return to MicronView for service.
Power button failure.
Return to MicronView for service.
11
Battery cannot be charged
when AC adapter is
working properly.
Battery is in poor contact.
Check and plug in the battery again.
Battery failure.
Contact the factory to replace the
battery
12
Battery cannot be fully
charged when AC adapter
supply.
Battery aging.
Contact the factory to replace the
battery
Internal parts failure.
Return to MicronView for service.
13
Device suddenly power
off when AC adapter is
connected.
AC adapter failure.
If the problem persists after replacing
another AC adapter, return to
MicronView for service.
14
Device suddenly powers
off when battery is
connected.
Low battery
Plug in the AC adapter and restart the
device.
Battery failure.
Contact the factory to replace the
battery
15
Flow error
Operating environment does not
meet requirements.
Operate the device in the required
environment.
Inlet is blocked.
Check if there is any blockage inside
the inlet. If not, Return to MicronView
or the local distributor for service.
Internal gas path failure.
Return to MicronView or the local
distributor for service.