User's Manual
23
Title
Description
Zone
Touch【Zone】icon to open the room list, and select the sampling room.
Status
Touch【Status】icon to select the status of the sampling room. The options are Operational,
At-Rest and As-Built.
Sampling SCM
Touch【Sampling SCM】icon to open the sampling scheme list, and select the sampling scheme.
Sampling Rule
Sampling rule consists of NORM mode and Custom mode. The NORM mode is used for the
samplings per ISO/EUGMP/CHINESEGMP/Fed Std. Under this mode, parameters cannot be
modified after the room and sampling scheme are selected (the background of sampling
conditions turn to grey.) If the sampling process is stopped, the result of the standard report is
invalid. The Custom mode is used for sampling for other purposes. Under this mode, sampling
parameters such as location, volume, sample time can be modified and a sampling can be
stopped at any moment.
Locations
The location cannot be changed under the NORM mode. It can be a selection list option under
the Custom mode.
Delay/
Interval
In auto mode, the first sample of each location complies with the delay. The following samples
comply with the interval. In the case of manual mode, all samples comply with the delay. The
sampling can be stopped during delay and interval. Press the start icon again to continue the
sampling. In the case of custom mode, delay and interval can be changed at any moment. The
countdown mode is used for delay and interval.
Volume
This shows the current sampled volume (above the line) and total volume of the sample (below
the line). In NORM mode, the volume cannot be changed.
Sample Time
This shows the current sampled time (above the line) and total time of the sample (below the
line). In NORM mode, the sample time cannot be changed.
Samples
This shows the number of samples which are being taken in the current location (left of the
split) and the total number of the samples (right of the split).
Operator
The operator’s account.
Beep
BEEP Mode: a beep will emit when sampling value of preset particle size range reaches the
threshold. Touch【Beep】icon to select the beep mode, and the parameter menu will display.
See chart 5-4-2. Set the particle type, particle range and threshold.
Auto Sample
If checked, the sampling will stop after all the samples on one location are finished.
Auto Print
If checked, the data will be printed automatically after each sampling is complete. If NORM
mode is selected, data of one location will be printed when sampling(s) of this location are all
completed. If manual mode is selected, data of each sample will be printed when this sample is
completed.
Graph
The curves show the instantaneous data of the 0.5μm particle, 5.0μm particle and bioparticle
channel in real time.
Location
display
It shows the number of locations in the sampling room. Flashing green light indicates the
sampling is underway in this location. Green light indicates the location which the sampling has
completed. Flashing red light indicates alarm appears during the sampling, and this location
should appear in red. In NORM mode, the location will appear in red light if sampling is
artificially terminated. See chart 5-4-3.
Set the counting mode, counting unit and size channel for the data displaying during the
sampling.
Counting mode consists of ∑ cumulative count and Δ differential count. Cumulative count: the