Chart 5-3-6 Save options
5.3.2 Sampling SCM
Sampling SCM is used for setting up the scheme for sampling. The scheme consists of volume, sample time, delay,
etc. It can be added, amended or removed in this menu. The sampling SCM can be associated with zoom here.
TouchSamplingicon of the left menu in the setting interface to set the scheme. See chart 5-3-7.
Chart 5-3-7 Sampling SCM
Title
Description
Name
The name of sampling SCM. Touch the input area to enter a new scheme, touch icon to
save the settings. Touch icon to select an existing scheme from the pull-down menu. Touch
icon to find a scheme, and touchOKicon to confirm. The scheme named default is a
system default sampling scheme. It can be modified, but it cannot be deleted. See chart 5-3-8.
Volume
The volume of air collected in each sampling.
Sample Time
The time of each sampling.
Delay
The time period before the first sampling is being taken.
Samples
The number of samples on one location.
Interval
The time period between the samples on one location.
Pumping Mode
Select Auto or Manual from the pull-down menu.
Auto: on one location, the sampling will stop after all the samples are finished.
Manual: on one location, the sampling will continue after pressing the start icon, when one