User's Manual

12
5.3.1 Zone
Zone is used for setting up and managing the room where the sample is being taken. The information can be
added, amended or removed in this menu. The sampling scheme can be associated with the room here. Touch
Zoneicon of the left menu in the setting interface to set up the room information. See chart 5-3-2.
Chart 5-3-2 Zone
Title
Description
Name
The name of the room where the sample is being taken. Touch the input area to enter a new
room and add room information, touch icon to save the settings. Touch icon to
select an existing room from the pull-down menu. Touch icon to find a room, and touch
OKicon to confirm. The room default is a system default room. It can be modified, but it
cannot be deleted. See chart 5-3-3.
Area
The area of the sampling room.
Standards
Select the standard from the pull-down menu. The standards include ISO 14644-1: 1999, ISO
14644-1: 2015, EUGMP-ISO: 1999, EUGMP-ISO: 2015, Chinese GMP 2010-ISO: 1999, Chinese
GMP 2010-ISO: 2015, Fed Std 209E F, Fed Std 209E M.
Locations
The number of locations in the sampling room. If the area is entered, locations will be
calculated automatically depending on the selected standard. The number can be modified.
The minimum value is 1.
Class
Select the class from the pull-down menu. If the standard is changed, the class will change,
accordingly.
Air Flow
The direction of the air flow in the sampling room. The options are Non-unidirectional and
Unidirectional.
Status
The status of the sampling room. The options are Operational, At-Rest and As-Built.
Largest selected
size
The size channel, according to the standard, is used to calculate the minimum sampling
volume.
Channels
Definition
Turn on/off the display and alert of the size channel and set the threshold of the alert value.
See chart 5-3-4.
Sampling SCM
Select an existing sampling scheme from the pull-down menu, or touch icon to add a new