Instruction Manual
Table Of Contents
- 1. Important Safety Instructions
- 2. Preparation and usage of this device
- 3. Cleaning and disinfecting
- 4. Maintenance, Care and Service
- 5. Malfunctions and Actions to take
- 6. Guarantee
- 7. Technical Specifications
- 1. Instrucciones importantes de seguridad
- 2. Preparación y uso del dispositivo
- 3. Limpieza y desinfección
- 4. Mantenimiento, cuidado y servicio posventa
- 5. Mensajes de error y soluciones
- 6. Garantía
- 7. Especificaciones técnicas
- 1. Importantes précautions d'emploi
- 2. Préparation et utilisation de l'appareil
- 3. Nettoyage et désinfection
- 4. Maintenance, entretien et service après-vente
- 5. Dépannage et actions à prendre
- 6. Garantie
- 7. Caractéristiques techniques
- 1. Importanti misure precauzionali
- 2. Preparazione e utilizzo del dispositivo
- 3. Pulizia e disinfezione
- 4. Manutenzione, conservazione e servizio
- 5. Malfunzionamenti e azioni da intraprendere
- 6. Garanzia
- 7. Specifiche tecniche
- 1. Sicherheitshinweise
- 2. Vorbereitung und Anwendung des Geräts
- 3. Reinigung und Desinfektion
- 4. Wartung, Pflege und Service
- 5. Fehlfunktionen und Massnahmen
- 6. Garantie
- 7. Technische Daten
- 1. Instruções de segurança importantes
- 2. Preparação e utilização do dispositivo
- 3. Limpeza e desinfeção
- 4. Manutenção, cuidados e serviço pós-venda
- 5. Falhas de funcionamento e soluções
- 6. Garantia
- 7. Especificações técnicas
- 1. Belangrijke veiligheidsinstructies
- 2. Voorbereiding en Gebruik
- 3. Reinigen en desinfecteren
- 4. Onderhoud, reinigen en vervangen
- 5. Problemen en mogelijke oplossingen
- 6. Garantie
- 7. Technische specificaties
- 1. Σημαντικές οδηγίες ασφαλείας
- 2. Προετοιμασία και χρήση αυτής της συσκευής
- 3. Καθαρισμός και απολύμανση
- 4. Συντήρηση, Φροντίδα και service
- 5. Δυσλειτουργία και Αντιμετώπιση
- 6. Εγγύηση
- 7. Τεχνικά χαρακτηριστικά
- 1. Önemli Güvenlik Talimatları
- 2. Aygıtın Hazırlanması ve Kullanılması
- 3. Temizlik ve Dezenfektasyon
- 4. Bakım, Onarım ve Servis
- 5. Arızalar ve Yapılacak İşlemler
- 6. Garanti Kapsamı
- 7. Teknik Özellikler
- IB NEB NANO basic AR-FA 1223.pdf
- 1. تعليمات الأمان الهامة
- 2. إعداد واستخدام الجهاز
- 3. التنظيف والتعقيم
- 4. الصيانة، والعناية، والخدمة
- 5. الأعطال، والإجراءات التي ينبغي إتخاذها
- 6. الضمان
- 7. المواصفات الفنية
- 1. نکات مهم درباره ایمنی
- 2. آمادگی و استفاده از این دستگاه
- 3. تمیز و ضدعفونی کردن
- 4. نگهداری، مراقبت و خدمات
- 5. عدم کارکرد صحیح و اقدامات لازم
- 6. ضمانت
- 7. مشخصات فنی
3Microlife NEB Nano basic
EN
1. Assemble the nebuliser kit AP. Ensure that all parts are
complete.
2. Fill the nebuliser with the inhalation solution as per your doctor’s
instructions. Ensure that you do not exceed the maximum level.
3. Connect one end of the air hose 8 to the dedicated outlet at
the bottom of the nebuliser AP and the other end to the air outlet
6 on the device.
4. Start the nebulisation by pressing the ON/OFF button 2; the
power indicator 3 will light up orange.
5. Place the mouthpiece 9 in the mouth or fit one of the face
masks over mouth and nose.
The mouthpiece gives you a better drug delivery to the
lungs.
Choose between adult AK or child face mask AL and make
sure that it encloses the mouth and nose area completely.
Use all accessories including the nose piece AT as
prescribed by your doctor.
6. During inhalation, sit upright and relaxed at a table and not in an
armchair, in order to avoid compressing your respiratory
airways and impairing the treatment effectiveness. Do not lie
down while inhaling. Stop inhalation if you feel unwell.
7. After completing the inhalation period recommended by your
doctor, press the ON/OFF button 2 to turn off the device.
8. Empty the remaining medication from the nebuliser and clean
the device as described in the section »Cleaning and disin-
fecting«
3. Cleaning and disinfecting
Thoroughly clean all components to remove medication residuals
and possible impurities after each treatment.
Use a soft and dry cloth with non-abrasive cleaners to clean the
compressor.
Cleaning and disinfecting of the accessories
Follow carefully the cleaning and disinfecting instructions of the
accessories as they are very important to the performance of the
device and success of the therapy.
Before and after each treatment
Disassemble the nebuliser 7 by turning the top counterclockwise
and remove the medicine conduction cone. Wash the components
of the disassembled nebuliser, the mouthpiece 9 and the nose
piece AT by using tap water; dip in boiling water for 5 minutes.
Reassemble the nebuliser components and connect it to the air
tube connector, switch the device on and let it work for 10-15
minutes.
4. Maintenance, Care and Service
Order all spare parts from your dealer or pharmacist, or contact
Microlife-Service (see foreword).
Replacement of the nebuliser
Replace the nebuliser 7 after a long period of inactivity, in cases
where it shows deformities, breakage, or when the vaporiser head
7-a is obstructed by dry medicine, dust, etc. We recommend to
replace the nebuliser after a period between 6 months and 1 year
depending on the usage.
Replacement of the air filter
In normal conditions of use, the air filter AM must be replaced
approximately after 200 working hours or after each year. We
recommend to periodically check the air filter (10 - 12 treatments)
and if the filter shows a grey or brown colour or is wet, replace it.
Extract the filter and replace it with a new one.
This device was designed for intermittent use of 30
min. On / 30 min. Off. Switch off the device after 30 min.
use and wait for another 30 min. before you resume treat-
ment.
The device requires no calibration.
No modification to the device is permitted.
Make sure that the internal parts of the device are not in
contact with liquids and that the power plug is discon-
nected.
Wash masks and air tube with warm water.
Only use cold disinfecting liquids following the manufac-
turer’s instructions.
Do not boil nor autoclave the air tube and masks.
Only use original nebulisers!