Microlife PT 200 1 2 PT200 Bluetooth® Patch Thermometer 3 Distributed by: Microlife USA, Inc. 1617 Gulf to Bay Blvd.,Floor 2 Clearwater, FL 33755,UNITED STATES Toll Free Help Line:1-800-568-4147 www.microlifeusa.
Microlife Bluetooth® Patch Thermometer PT 200 Microlife PT 200 1 2 3 4 5 6 7 5 1 + Measuring sensor ON/OFF button Power indicator Battery Compartment How to insert the battery Preparing the device for use Correct positioning of the device Read the instructions carefully before using this device. Type BF applied part 6 This Microlife thermometer is a high quality product incorporating the latest technology and tested in accordance with international standards.
the displayed circle will turn red (high temperature) or blue (low temperature) and an alarm beep will be triggered. Please consult with your doctor if you do not know the appropriate body temperature. Bluetooth® Function This device connects to the «Microlife Thermo 24» App by using Bluetooth® 4.0 and enables easy monitoring of temperature. 2. Important Safety Instructions Follow instructions for use. This document provides important product operation and safety information regarding this device.
3. Take one adhesive patch, remove the protective film and place the device right in the middle of the patch 6. The battery compartment 4 side should be stuck on the patch. The measuring sensor 1 needs to be in contact with the skin and as close as possible to the armpit. 4. Place the device with the adhesive patch directly on the skin below the armpit 7. 5. Lower the arm and wait around 15 minutes until the temperature rises to the body temperature level. 6.
9. Technical Specifications Type: Patch Thermometer PT 200 Measurement Operation: Direct Mode Measurement range: 32.0 - 43.0 °C / 89.6 - 109.4 °F Resolution: Measurement accuracy: Memory: Operating conditions: Storage conditions: Automatic Switch-off: Communication: Compatibility: Battery: Battery lifetime: Dimensions: Weight: IP Class: Reference to standards: Expected service life: 0.1 °C / °F Laboratory: ± 0.1 °C / ± 0.
Appendix Guidance and manufacturer’s declaration – electromagnetic emissions The model PT 200 is intended for use in the electromagnetic environment specified below. The customer or the user of the model PT 200 should assure that it is used in such an environment. Emissions RF emissions CISPR 11 Compliance Group 1 Electromagnetic environment-- guidance The model PT 200 uses RF energy only for its internal function.
1. Guidance and manufacturer’s declaration – electromagnetic immunity The model PT 200 is intended for use in the electromagnetic environment specified below. The customer or the user of the model PT 200 should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment --guidance Electrostatic discharge (ESD) IEC 61000-4-2 ±8 kV contact ±15 kV air ±8 kV contact ±15 kV air Floors should be wood, concrete or ceramic tile.
2. Guidance and manufacturer’s declaration – electromagnetic immunity The model PT 200 is intended for use in the electromagnetic environment specified below. The customer or the user of the model PT 200 should assure that it is used in such an environment.
3. Recommended separation distances between portable and mobile RF communications equipment and the model PT 200 The model PT 200 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.