Microlife FR1MF1-B(NC150 BT) 1 2a FR1MF1-B (NC150 BT) Non Contact Thermometer EN Distributed by: Microlife USA, Inc. 1617 Gulf to Bay Blvd.,Floor 2 Clearwater, FL 33755,UNITED STATES Toll Free Help Line:1-800-568-4147 www.microlifeusa.
Microlife Non Contact Thermometer FR1MF1-B(NC150 BT) Microlife FR1MF1-B(NC150 BT) 2 1 1 2 3 4 5 6 7 8 9 AT AK AL AM AN AO AP AQ AR AS BT BK BL BM BN BO BP BQ BR 3 4 6 M 8 7 5 9 AP BM AT AQ BN AK AR BO AL AS BP AM BT 2 1 3 Measuring sensor Tracking light Display START button ON/OFF button Mode switch Battery compartment cover M-button (memory) All segments displayed Memory Ready for measuring Measurement complete Body mode Object mode Low battery indicator Changing between Celsius and F
13. Battery Replacement 14. Guarantee 15. Technical Specifications 16. www.microlife.com Guarantee Card (see Back Cover) Bluetooth® Function This device connects to the «Microlife Connected Health+» App by using Bluetooth® 4.0 and enables easy monitoring of temperature. 1. The Advantages of this Thermometer Follow instructions for use. This document provides important product operation and safety information regarding this device.
The measurement results given by this device WARNING: is not a diagnosis. It is not replacing the need for the consultation of a physician, especially if not matching the patient’s symptoms. Do not rely on the measurement result only, always consider other potentially occurring symptoms and the patient’s feedback. Calling a doctor or an ambulance is advised if needed. 3. How this Thermometer measures Temperature This thermometer measures infrared energy radiated from the forehead as well as objects.
NOTE: Patients and thermometer should stay in similar room condition for at least 30 minutes. Don't take a measurement while or immediately after nursing a baby. Don't use the thermometer in high humidity environments. Patients should not drink, eat or exercise before/while taking the measurement. Don't move the measurement device from the measuring area before hearing the termination beep.
The Bluetooth® icon indicator AK flashes, when connecting with a device or uploading data to the device. The Bluetooth® function is automatically switched off during measurement; after measurement the device will switch it on and upload the data to the «Microlife Connected Health+» App. App Tutorial To access the tutorial, select «Tutorial» in the menu of the «Microlife Connected Health+» App (located in the upper left corner on the screen). 11.
15. Technical Specifications Non Contact Thermometer Type: Measurement Body mode: 34.0 - 43.0 °C / 93.2 - 109.4 °F Object mode: 0.1 - 99.9 °C / 32.2 - 211.8 °F range: Resolution: 0.1 °C / °F Measurement Body mode: accuracy 0.2 °C, 35.0 42.0 °C / 0.4 °F, 95.0 107.6 °F (Laboratory): 0.3 °C, 34.0 34.9 °C and 42.1 43.0 °C / 0.5 °F, 93.2 ~ 94.8 °F and 107.8 ~ 109.4 °F Object mode: ±1.0 °C, 0.1 ~ 99.9 °C / ±2 °F, 32.2 ~ 211.
This device complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. Electromagnetic compatibility: This device fulfills the stipulations of the standard IEC 60601-1-2 (EMC). Technical alterations reserved.
Appendix Guidance and manufacturer’s declaration – electromagnetic emissions The model FR1MF1-B(NC150 BT) is intended for use in the electromagnetic environment specified below. The customer or the user of the model FR1MF1-B(NC150BT) should assure that it is used in such an environment. Emissions RF emissions CISPR 11 Compliance Group 1 Electromagnetic environment-- guidance The model FR1MF1-B(NC150 BT) uses RF energy only for its internal function.
1. Guidance and manufacturer’s declaration – electromagnetic immunity The model FR1MF1-B(NC150 BT) is intended for use in the electromagnetic environment specified below. The customer or the user of the model FR1MF1-B(NC150 BT) should assure that it is used in such an environment.
2. Guidance and manufacturer’s declaration – electromagnetic immunity The model FR1MF1-B(NC150 BT) is intended for use in the electromagnetic environment specified below. The customer or the user of the model FR1MF1-B(NC150 BT) should assure that it is used in such an environment.
3. Recommended separation distances between portable and mobile RF communications equipment and the model FR1MF1-B(NC150 BT) The model FR1MF1-B(NC150 BT) is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.