Instruction Manual

EN
23
WatchBP Office
Operation temperature/
humidity:
10 to 40 °C (50 to 104 °F)/ 15 - 90 % relative maximum humidity
Storage temperature/
humidity:
-20 to 55 °C (-4 to 131 °F)/ 15 - 90 % relative maximum humidity
Weight:
620g (including rechargeable battery pack)
Dimensions:
220.4 x 121.7 x 63.3 mm
Measuring method:
Oscillometric, Systolic blood pressure = K1; Diastolic blood pressure
= K5
Measurement range:
60 - 255mmHg - systolic blood pressure; 30 - 200mmHg - diastolic
blood pressure; 30 - 239 beats per minute - pulse
Cuff pressure display:
Range: 0 - 299 mmHg; Resolution: 1 mmHg; Static accuracy:
pressure within ± 3 mmHg;
Pulse accuracy:
±5 % of the readout value
Power source:
Rechargeable battery pack; 4.8V 2400 mAh; Mains power supply DC
7.5V, 1.5 A
Expected service life:
2 years
Reference to
Standards:
Device corresponds to the requirements of the standard for non-
invasive blood pressure monitor.
IEC 60601-1: 2005+A1:2012
IEC 60601-1-2:2014
ANSI/AAMI/ISO 81060-2
ANSI/AAMI/IEC 80601-2-30
Electromagnetic
Compatibility:
Device fullls the stipulations of the standard IEC 60601-1-2.
The stipulations of the EU Directive 93/42/EEC for Medical Devices
Class IIa have been fullled.
Type BF applied part
Microlife reserves the right to alter technical specications without prior
written notice.
This device is covered by a two-year guarantee from the date
of purchase. This guarantee is valid only on presentation of the
guarantee card completed by the owner conrming date of purchase
or purchase receipt. Batteries and wearing parts are not covered by
this guarantee.