Instruction Manual
EN
23
WatchBP Office
Operation temperature/
humidity:
• 10 to 40 °C (50 to 104 °F)/ 15 - 90 % relative maximum humidity
Storage temperature/
humidity:
• -20 to 55 °C (-4 to 131 °F)/ 15 - 90 % relative maximum humidity
Weight:
• 620g (including rechargeable battery pack)
Dimensions:
• 220.4 x 121.7 x 63.3 mm
Measuring method:
• Oscillometric, Systolic blood pressure = K1; Diastolic blood pressure
= K5
Measurement range:
• 60 - 255mmHg - systolic blood pressure; 30 - 200mmHg - diastolic
blood pressure; 30 - 239 beats per minute - pulse
Cuff pressure display:
• Range: 0 - 299 mmHg; Resolution: 1 mmHg; Static accuracy:
pressure within ± 3 mmHg;
Pulse accuracy:
• ±5 % of the readout value
Power source:
• Rechargeable battery pack; 4.8V 2400 mAh; Mains power supply DC
7.5V, 1.5 A
Expected service life:
• 2 years
Reference to
Standards:
• Device corresponds to the requirements of the standard for non-
invasive blood pressure monitor.
IEC 60601-1: 2005+A1:2012
IEC 60601-1-2:2014
ANSI/AAMI/ISO 81060-2
ANSI/AAMI/IEC 80601-2-30
Electromagnetic
Compatibility:
• Device fullls the stipulations of the standard IEC 60601-1-2.
The stipulations of the EU Directive 93/42/EEC for Medical Devices
Class IIa have been fullled.
Type BF applied part
Microlife reserves the right to alter technical specications without prior
written notice.
This device is covered by a two-year guarantee from the date
of purchase. This guarantee is valid only on presentation of the
guarantee card completed by the owner conrming date of purchase
or purchase receipt. Batteries and wearing parts are not covered by
this guarantee.
Product: WatchBP Office
Product number: BP3SK1-3B
Date:
Name:
Address:
Date:
Telephone:
Email:
Guarantee Card