Instruction Manual
EN
25
WatchBP Office
Guidance and MANUFACTURER’S declaration – electromagnetic
IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic immunity
The BP3SK1-3B is intended for use in the electromagnetic environment specied below. The
customer or the user of the BP3SK1-3B should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
6 Vrms 150 kHz
to 80 MHz
outside
ISM bandsa
3V/m
80 MHz to 2.7
GHz
3 Vrms
150 kHz
to 80
MHz
6 Vrms 150
kHz to 80
MHz
outside
ISM bandsa
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Models BP3SK1-3B, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres(m).
Field strengths from xed RF transmitters,
as determined by an electromagnetic
site survey, a should be less than the
compliance level in each frequency range b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reection from structures, objects and people.