Guidance and manufacturer´s declaration electromagnetic immunity
The Model BP3SK1-3B are intended for use in the electromagnetic environment specied below.
The customer or the user of the Model BP3SK1-3B should assure that it is used in such an
environment.
Immunity test IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
± 8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
± 8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Floors should be wood, concrete
or ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
100 kHz repetition
frequency ± 1 kV
for input/output
lines
± 2 kV for power
supply lines 100
kHz repetition
frequency
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC
61000-4-5
± 0.5 kV, ± 1 kV
differential mode
line-line
± 0.5 kV, ± 1 kV
differential mode
line-line
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
0 % UT (100 % dip
in UT ) for 0.5 cycle
at 0°, 45°, 90°,
135°,180°, 225°,
270°, and 315°
0 % UT (100 % dip
in UT )
for 1 cycle at
70 % UT (30 % dip
in UT ) for 25/30
cycles at
0 % UT (100 % dip
in UT ) for 250/300
cycle at
0 % UT (100 % dip
in UT ) for 0.5 cycle
at 0°, 45°, 90°,
135°,180°, 225°,
270°, and 315°
0 % UT (100 % dip
in UT )
for 1 cycle at
70 % UT (30 % dip
in UT ) for 25/30
cycles at
0 % UT (100 % dip
in UT ) for 250/300
cycle at
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the models BP3SK1-3B product
name requires continued operation
during power mains interruptions,
it is recommended that the models
BP3SK1-3B be powered from an
uninterruptible power supply or a
battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-
4-8
30 A/m, 50/60Hz 30 A/m, 50/60Hz
Power frequency magnetic
elds should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity – for all