Instruction Manual
EN
23
WatchBP Office
Operation temperature/
humidity:
• 10 to 40 °C (50 to 104 °F)/ 15 - 90 % relative maximum humidity
Storage temperature/
humidity:
• -20 to 55 °C (-4 to 131 °F)/ 15 - 90 % relative maximum humidity
Weight:
• 620g (including rechargeable battery pack)
Dimensions:
• 220.4 x 121.7 x 63.3 mm
Measuring method:
• Oscillometric, Systolic blood pressure = K1; Diastolic blood pressure
= K5
Measurement range:
• SYS/DIA:30-280mmHg
Pulse: 40-200 per minute
Cuff pressure display:
• Range: 0 - 299 mmHg; Resolution: 1 mmHg; Static accuracy:
pressure within ± 3 mmHg;
Pulse accuracy:
• ±5 % of the readout value
Power source:
• Rechargeable battery pack; 4.8V 2400 mAh; Mains power supply DC
7.5V, 1.5 A
Expected service life:
• 2 years
Reference to
Standards:
• Device corresponds to the requirements of the standard for non-
invasive blood pressure monitor.
IEC 60601-1
IEC 60601-1-2
ANSI/AAMI/ISO 81060-2
ANSI/AAMI/IEC 80601-2-30
Electromagnetic
Compatibility:
• Device fullls the stipulations of the standard IEC 60601-1-2.
Type BF applied part
Microlife reserves the right to alter technical specications without prior
written notice.
1
Guidance and manufacturer´s declaration – electromagnetic emission
2
The model BP3SK1-3B is intended for use in the electromagnetic environment specied
below. The customer or the user of the model BP3SK1-3B should assure that it is used in
such an environment.
3
Emissions test
Compliance
Electromagnetic environment – guidance
4
RF emissions
CISPR 11
Group 1 The Model BP3SK1-3B uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
5
RF emissions
CISPR 11
Class A The BP3SK1-3B is suitable for use in all
establishments, other than domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
6
Harmonic
emissions IEC
61000-3-2
Not
applicable
7
Voltage
fluctuations /
flicker emissions
IEC 61000-3-3
Not
applicable
Annex
Guidance and manufacturer´s declaration – electromagnetic emission – for all
EQUIPMENT AND SYSTEMS
Row