WatchBP Office Professional Automated Office Blood Pressure Monitor BP 3SK1-3B Instruction Manual
Federal Communications Commission (FCC) Statement changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Table of Contents WatchBP Office Product description ・Contents ................................................................................................................................ 6 ・Model Type . ........................................................................................................................... 6 ・Upgrading the version of the device.......................................................................................... 6 ・Product Overview .............................
Product description The WatchBP Office consists of two major parts • The device, cuffs and accessories. • The WatchBP Analyzer Software. With the WatchBP Analyzer Software 1) The device can be programmed for the blood pressure measurement procedure. 2) Measured blood pressure values can be downloaded to the PC. 3) A PDF report and Microsoft Excel spreadsheet for data analysis can be generated. the latest WatchBP Analyzer Software from the Microlife website. * Download https://www.microlife.
WatchBP Office Product Overview Cuff Socket Power Adapter Socket Display USB port DC 7.5V 1.
Initial set up Attaching the power plug to the power adapter Select a suitable plug attachment and attach to the power adapter as shown here. Charge the battery completely When using the device for the first time, charge the battery until the recharge indicator on the device turns green. Power ON/OFF Press button to switch on the device. Press and hold button for 3 seconds to switch off the device and turn off the LCD screen. The device displays ‘oFF’ before turning off.
Before using the device WatchBP Office Selecting the correct cuff A variety of different cuff sizes are available. M and L size cuffs are provided with the device. Use the cuff marker to select the cuff size that best matches the circumference of the patient’s upper arm. Cuff Size Circumference (cm) Circumference (inch) S 14-22 5.5-8.7 M 22-32 8.7-12.6 L 32-42 12.6-16.5 L-XL 32-52 12.6-20.5 * Each cuff is provided with 130 cm air tube.
Taking measurements in MANUAL and AUTO Mode Turn on the power Turn on the device by pressing the of the device. button Connect the cuff to the device Connect the cuff to the device by inserting the cuff connector into the cuff connector socket. Select an operation mode There are two measurement modes that can be used. Press the button to switch between AUTO or MANUAL Mode.
WatchBP Office 4 Set the Atrial fibrillation (AFIB) detector – Press the or button to switch the AFIB detector ON or OFF. Press button to confirm. the AFIB detector option appears only for the device * Set version with AFIB detector. If the device has an activated AFIB detector then it is witched on in default. 5 Set the Central blood pressure (CBP) measurement – Press or button to switch ON or OFF CBP measurement. Press button to confirm.
Taking measurement in AUTO Mode Select AUTO mode. Press the button to perform automatic measurements based on the settings of AUTO mode. The device shows all the settings and then starts counting down the Resting Time before the first measurement. The average measurement reading is displayed and saved after the measurements are complete.
WatchBP Office Hide Cuff Pressure during deflation – This option helps you to determine Korotkoff K1 and K5 sound readings without digit preference. To use the HIDE function in MANUAL mode, push button to select the setting of HIDE function and confirm the selection by button and finish the setting of MANUAL mode. Taking measurement in MANUAL mode Use a stethoscope to perform measurement.
Special Functions Screening for atrial fibrillation during blood pressure measurement The device is designed to screen for atrial fibrillation during blood pressure measurements (optional) with high accuracy: a sensitivity of 98% and a specificity value of 92%*. If atrial fibrillation is detected this will be shown in the report. et al. Screening for atrial fibrillation with automated blood pressure measurement: Research evidence and F Verberk practice recommendations. Int J Cardiol 2016: 465–473.
WatchBP Office The time that is needed to determine the central blood pressure value may vary among patients; i.e. with faster heart rate, less time is required for collecting the number of required PVP waveforms. It is very important to keep the arm still during the time the PVP waveforms are collected.
Using WatchBP Analyzer The memory data can be transferred to your PC (personal computer) running the WatchBP Analyzer Software by connecting the monitor via USB cable or Bluetooth. System Requirements for Software: 1GHz CPU. 512MB Memory, 4.5GB free hard disk space, Microsoft Windows 7 SP1 / 8 / 10 Installing the Software Program The latest WatchBP Analyzer Software is available from the Microlife website. https://www.microlife.
Bluetooth connectivity WatchBP Office Bluetooth connection of the WatchBP Analyzer supports Microsoft Windows 10 Regarding the connectivity architecture design, we used proprietary communication protocol to do data transfer process. The specific communication protocol is assured that the information (data) is correct. The program checks ACK firstly. Afterward, the program compares the received checksum with the sum of encoded raw data. If the result is correct, the data is guaranteed during transmission.
Rechargeable battery and power adapter Rechargeable Battery The device has a built-in, rechargeable Ni-MH battery pack that can perform up to 400 measurement cycles on a full charge. The battery can be recharged with the power adapter provided with the device. The empty battery indicator is displayed when the battery is low. • When using the device for the first time, charge the battery until the recharge indicator turns to green. • The orange recharge indicator indicates that the recharge is in progress.
WatchBP Office • Only activate the pump when the cuff is connected to the device. • Do not use the device if you think it is damaged or if anything appears unusual. • Read the further safety instructions in the individual sections of the instruction manual. Observe the storage and operating conditions as described in the “Technical specifications” section of this manual.
Error messages and Troubleshooting If an error occurs during measurement, the measurement is interrupted and an error message «Er» is displayed. Error Description Potential cause and remedy "Er 1" Signal too The pulse signals on the cuff are too weak. Reposition the cuff and repeat the measurement. weak "Er 2" Error signal During the measurement, error signals were detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.
WatchBP Office "Er 15" Abnormal result of central blood The measuring signals are inaccurate so that no result can be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement. pressure reading "Er F" The device has gone into “single fault condition” “Er A” "HI" Flash memory Possible hardware fault. Try again. If the error persists, contact error Microlife or the local distributor.
Technical specifications Operation temperature/ • 10 to 40 °C (50 to 104 °F)/ 15 - 90 % relative maximum humidity humidity: Storage temperature/ humidity: • -20 to 55 °C (-4 to 131 °F)/ 15 - 90 % relative maximum humidity Weight: • 620g (including rechargeable battery pack) Dimensions: • 220.4 x 121.7 x 63.
WatchBP Office Annex Guidance and manufacturer´s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS Row 1 Guidance and manufacturer´s declaration – electromagnetic emission 2 The model BP3SK1-3B is intended for use in the electromagnetic environment specified below. The customer or the user of the model BP3SK1-3B should assure that it is used in such an environment.
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS Guidance and manufacturer´s declaration – electromagnetic immunity The Model BP3SK1-3B are intended for use in the electromagnetic environment specified below. The customer or the user of the Model BP3SK1-3B should assure that it is used in such an environment.
WatchBP Office Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY Guidance and manufacturer’s declaration – electromagnetic immunity The BP3SK1-3B is intended for use in the electromagnetic environment specified below. The customer or the user of the BP3SK1-3B should assure that it is used in such an environment.
a The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
WatchBP Office Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM Recommended separation distances between portable and mobile RF communications equipment and the model BP3SK1-3B The Model BP3SK1-3B is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Recommended separation distances between RF wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between RF wireless communications equipment and the device as recommended below, according to the maximum output power of the communications equipment. Maximum Power W Distance 385 1.8 0.
WatchBP Office Guarantee Card This device is covered by a two-year guarantee from the date of purchase. This guarantee is valid only on presentation of the guarantee card completed by the owner confirming date of purchase or purchase receipt. Batteries and wearing parts are not covered by this guarantee.
Distributed by: Microlife USA, Inc. 1617 Gulf to Bay Blvd 2nd Floor, Suite A Clearwater, FL 33755, USA Tel. +1 727 442 5353 Fax +1 727 442 5377 Email: msa@microlifeusa.com www.watchbp.