Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Model BP3MX1-3C(WatchBP Home A BT) are intended for use in the electromagnetic
environment s
pecified below. The customer or the user of the Model BP3MX1-3C(WatchBP Home A
BT) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
±8kVair
± 6 kV contact
±8kVair
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
±1kVfor
input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
±2kVcommon
mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations
on power supply
input lines
IEC 61000-4-11
<5%UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60%dipinUT)
for 5 cycles
70 % UT
(30%dipinUT)
for 25 cycles
<5%UT
(>95 % dip in UT )
for 5 sec
<5%UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
<5%UT
(>95 % dip in UT )
for 5 sec
Mains power quality s
hould be that
of a typical commercial or hospital
environment. If the user of the
models BP3MX1-3C(WatchBP
Home A BT) product name
requires continued operation
during power mains interruptions, it
is recommended that the models
BP3MX1-3C(WatchBP Home A
BT) be powered from an
uninterruptible power supply or a
battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.