User's Manual

21
11. Technical specifications:
Weight: 400 g (with batteries)
Size: 3.5” (W) x 6” (L) x 1.5” (H)
Storage temperature: 20 to +55°C (-4°F to +131°F)
Humidity: 15 to 90% relative humidity maximum
Operation temperature: 10 to 4C ( 5F - 104°F)
Display: LCD-Display (Liquid Crystal Display)
Measuring method: oscillometric
Pressure sensor: capacitive
Measuring range:
SYS: 60 to 255 mmHg
DIA: 40 to 200 mmHg
Pulse: 40 to 199 per minute
Cuff pressure display range: 0–299 mmHg
Memory: Automatically stores the last 99
measurements for 2 users (total 198)
Measuring resolution: 1 mmHg
Accuracy: pressure within ± 3 mmHg
or 2% of reading > 200mmHg
pulse ± 5 % of the reading
Power source: a) 4 AA batteries, 1.5 V
b) AC adapter 6 V DC 600 mA
(voltage 4.5 V DC to 6 V DC)
Accessories: One size easy fit cuff 22-46 cm
(8.7” - 18.1”)
Technical alterations reserved!
Made in China
Read the instruction manual carefully before using this device, especially
the safety instructions, and keep the instruction manual for future use.
Type BF applied part.
*Not for use on children 12 years of age or younger.