User's Manual

BP3KN1-3B
22
IB BP3KN1-3B V4920 EN
EN
12. Technical specifications
Operating conditions: ………10 - 40 °C / 50 - 104 °F
15 - 90 % relative maximum humidity
Storage conditions: …………-20 - +55 °C / -4 - +131 °F
15 - 90 % relative maximum humidity
Weight: …………………………120 +/-5 g (including batteries)
Dimensions: ……………………82 x 64 x 13.9 mm
Cuff size: ………………………13.5 – 23cm (5.25 – 9.06 inches)
Measuring procedure: ………Oscillometric, corresponding to Korotkoff method: Phase I
systolic, Phase V diastolic
Measurement range: …………SYS: 60~255 mmHg
DIA: 40~200 mmHg
40 - 199 beats per minute – pulse
Cuff pressure display range: 0 - 299 mmHg
Resolution: ………………………1 mmHg
Static accuracy: ……………… pressure within ± 3 mmHg
Pulse accuracy: …………………± 5 % of the readout value
Voltage source: …………………2 x 1.5 V alkaline batteries; size AAA
Battery lifetime: …………………approx. 300 measurements (using new batteries)
IP Class: ………………………… IP22
Expected service life: ………… Device: 5 years or 10000 measurements
Accessories: 2 years
Atmospheric pressure range…70KPa -106KPa
Accessories included………user manualStorage BoxGift boxBP diary2xAAA
Batteries
Technical alterations reserved.