Product Info
With the unit off, press the «MEMORY» button. Pressing the «MEMORY» button again displays
the previous value. Pressing the «MEMORY» button repeatedly enables you to move from one
stored value to another. If an irregular heartbeat is detected, the symbol xxxxxx is displayed
including the time and the result of the measurement.
NOTE: Measurements for each user are averaged and stored separately. Be certain that you
are viewing the measurements for the correct user.
NOTE: Once the maximum memory capacity of 99 is exceeded, the oldest value is
automatically overwritten. Values should be evaluated by a doctor before the memory capacity
is reached, otherwise data will be lost.
Further information
Measurements should not occur soon after each other; otherwise, the results will be skewed.
Wait for several minutes in a relaxed position, sitting or lying, before you repeat a measurement.
Discontinuing a Measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g., the patient
feels unwell), the «ON/OFF» button can be pressed at any time. The device immediately lowers
the cuff pressure automatically
Delete Memory
Before you delete all memory readings stored, make sure you won’t need to refer to them at a
later date. Keeping a written record is prudent and may provide additional information for your
doctor’s visit.
To delete stored memory readings, hold the «MEMORY» button down until the LCD shows a
«CL»«ALL»». Release the button. Press the «MEMORY»or set?? button while «CL» is flashing
to complete the operation. It is impossible to erase the values individually. (show picture)
ERROR MESSAGES/TROUBLESHOOTING
If an error occurs during a measurement, the measurement is interrupted, and a corresponding
error code is displayed, example <Err 3> (cc has picture)
Error Description Possible
Cause(s)/Solutions
Cuff picture
?
?
Err 1 Signal too weak The pulse signals on the cuff
are too weak. Re-position
the cuff and repeat the
measurement.
Err 2 Error signal During the measurement,
error signals were detected
by the cuff, caused for
instance by movement or