for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1.
Plug the adapter cable into the mains adapter socket (4)
in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
12. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message,
e.g. «ERR 3», is displayed.
Error
Description
Potential cause and remedy
«ERR Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and
repeat the
measurement.*
«ERR
(26)
Error signal During the measurement, error signals
were detected by the cuff,
caused for
instance by movement or muscle
tension. Repeat the
measurement,
keeping your arm still.
«ERR
(23)
No pressure
in
the cuff
An adequate pressure cannot be generated in the cuff. A leak may
have
occurred. Check that the cuff is correctly
connected and is not
too loose. Replace
the batteries if necessary. Repeat the
«ERR
Abnormal
result
The measuring signals are inaccurate
and no result can therefore be
displayed.
Read through the checklist for
performing reliable
measurements and
then repeat the measurement.*
«ERR
AFIB/MAM
Mode
There were too many errors during the
measurement in AFIB/MAM
mode,
making it impossible to obtain a final
result. Read through the
checklist for
performing reliable measurements and
then repeat the
«HI» Pulse or cuff
pressure
too
The pressure in the cuff is too high (over
300 mmHg) OR the pulse
is too high
(over 200 beats per minute). Relax for 5
minutes and
«LO» Pulse too
low The pulse is too low (less than 40 beats
per minute). Repeat the
Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
13. Safety, Care, Accuracy Test and Disposal
Safety and protection
This device may only be used for the purposes described in these instructions. The manufacturer
cannot be held liable for
damage caused by incorrect application.
This device comprises sensitive components and must be
treated with caution. Observe the
storage and operating conditions described in the «Technical Specifications» section.
Protect it from: