User's Manual
Guidance and manufacturer’s declaration-electromagnetic immunity
The BP3GT1-6B is intended for use in the electromagnetic environment specified below.
The customer or the user of the BP3GT1-6B should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic
environment-guidance
Electrostatic
discharge(ESD) IEC
61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%
Electrical fast
transient/burst IEC
61000-4-4
+ 2kV for power
supply lines
+ 1kV for input/output
lines
+ 2kV for power
supply lines
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5 + 1kV line(s) to line(s)
+ 2kV line(s) to earth
+ 1kV differential
mode
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the BP3GT1-6B
requires continued operation
during power mains
interruptions, it is recommended
that the BP3GT1-6B
be powered
from an uninterruptible power
supply or a battery.
Power frequency(50, 60
Hz) magnetic field IEC
61000-4-8
3 A/m 3 A/m
The BP3GT1-6B power
frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.