Operation Manual
4
toring, 2001, 6:145-147» for oscillometric instruments). A count 
down indicates the remaining time.
 The individual results are not displayed. Your blood pressure 
will only be displayed after all 3 measurements are taken.
 Do not remove the cuff between measurements.
 If one of the individual measurements was questionable, a 
fourth one is automatically taken.
3. Taking a Blood Pressure Measurement using this 
Device
Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the 
measurement.
2. Sit down for at least 5 minutes before the measurement and relax.
3. Always measure on the same arm (normally left). It is recom-
mended that doctors perform double arm measurements on a 
patients first visit in order to determine which arm to measure in 
the future. The arm with the higher blood pressure should be 
measured.
4. Remove close-fitting garments from the upper arm. To avoid 
constriction, shirt sleeves should not be rolled up - they do not 
interfere with the cuff if they are laid flat.
5. Always ensure that the correct cuff size is used (marking on the 
cuff).
 Fit the cuff closely, but not too tight.
 Make sure that the cuff is positioned 2 cm above the elbow.
 The artery mark located on the cuff (ca. 3 cm long bar) must 
lie over the artery which runs down the inner side of the arm.
 Support your arm so it is relaxed.
 Ensure that the cuff is at the same height as your heart.
6. Press the ON/OFF button 1 to start the measurement.
7. The cuff will now pump up automatically. Relax, do not move 
and do not tense your arm muscles until the measurement 
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and the 
pressure falls gradually. If the required pressure was not reached, 
the device will automatically pump some more air into the cuff.
9. During the measurement, the pulse indicator BT flashes in the 
display.
10.The result, comprising the systolic AO and the diastolic AP 
blood pressure and the pulse rate AQ is displayed. Note also the 
explanations on further display symbols in this booklet.
11.When the device has finished measuring, remove the cuff.
12.Switch off the device. (The monitor does switch off automati-
cally after approx. 1 min.).
How not to store a reading
As soon as the reading is displayed press and hold the ON/OFF 
button 1 until «M» BO is flashing. Confirm to delete the reading by 
pressing the M-button AT.
You can stop the measurement at any time by pressing the 
ON/OFF button (e.g. if you feel uneasy or an unpleasant 
pressure sensation). 
If the systolic blood pressure is known to be very high
, 
it can be an advantage to set the pressure individually. Press 
the ON/OFF button after the monitor has been pumped up to 
a level of approx. 30 mmHg (shown on the display). Keep the 
button pressed until the pressure is about 40 mmHg above 
the expected systolic value – then release the button. 
4. Appearance of the Pulse Arrhythmia Indicator for 
early Detection 
This symbol BM indicates that certain pulse irregularities were 
detected during the measurement. In this case, the result may 
deviate from your normal blood pressure – repeat the measure-
ment. In most cases, this is no cause for concern. However, if the 
symbol appears on a regular basis (e.g. several times a week with 
measurements taken daily) we advise you to tell your doctor. 
Please show your doctor the following explanation:
5. Traffic Light Indicator in the Display
The bars on the left-hand edge of the traffic light display AR show 
you the range within which the indicated blood pressure value lies. 
Information for the doctor on frequent appearance of the 
Arrhythmia indicator
This device is an oscillometric blood pressure monitor that also 
analyses pulse irregularity during measurement. The device is 
clinically tested.
The arrhythmia symbol is displayed after the measurement, if 
pulse irregularities occur during measurement. If the symbol 
appears more frequently (e.g. several times per week on 
measurements performed daily) we recommend the patient to 
seek medical advice.
This device does not replace a cardiac examination, but serves 
to detect pulse irregularities at an early stage.










