User guide

Chapter 3 MicroLab Software
94
Temperature Calibration
Note: For the MicroLog model, the Temperature calibration
is performed in the software. The logger itself is not
calibrated. When calibrating the MicroLogPRO model,
calibration is performed directly onto the data logger,
meaning the logger itself is calibrated, not the software.
1. Connect the data logger to the PC.
2. Click Logger on the main menu, then click Calibration.
3. Enter the calibration password, then click OK.
4. Select the desired temperature sensor in the Choose
sensor drop-down menu.
5. Click Default to restore the original values.