6. Click Default to restore the original values.
7. Disconnect the data logger from the PC and insert it into
the humidity chamber.
8. Set the humidity chamber to the first reference value.
Wait until the humidity level is stabilized and write down
the logger’s reading.
9. Repeat the last step with the second reference value.
10. Connect the logger to the PC and enter the Calibration
dialog.
11. Enter the two logger values into the MicroLog value text
boxes.
12. Enter the two reference values into the Reference value
text boxes.
13. Click Calibrate.
To restore default calibration, click Default.