Technical data

MicroLab PC
Agilent MicroLab Software Operation Manual 23
Component Reporting page
Component Reporting enables the method developer to determine
when a component is shown in both the Results screen and the
report. Only selected components that are selected in the Display
column will be shown on the screen and the report.
Edit Component Reporting Condition dialog box
Additionally, the method developer can choose to report a
component only under certain conditions. For instance, the
developer can specify that a component is reported only when its
value is within the range for which it has been calibrated. This
enables calibrations over several ranges (high, medium and low) to
be present in a method. If specified, the software will only report the
concentration from the calibration that is in range. For multivariate
methods, any of the method statistical measurements (Mahalanobis
Distance, F-test, and so on) can be used to determine if a value
should be reported.
Report As: This section enables the developer to choose if the
component is reported as the calculated value or as a text string
based on that value. Used with the Boolean logic in the condition
setting, this allows text-based results, such as ‘out of specification’ or
‘good’, to be displayed instead of numeric results.
Component/Diagnostic: This drop-down list shows all of the
diagnostic information that can be used to make a display logic
choice.
Test State: Enables specification of the state of the diagnostic. This is
used with ‘Value’ below.