Technical data

MicroLab PC
Agilent MicroLab Software Operation Manual 19
Editing
Info page
Method Name: This field specifies the method name; the method
name will be displayed in the method list.
Developer: This field specifies the user who developed the method.
Comment: This field is made available to the user to input any
special information concerning the method.
Type page
Method Type: Three types of methods are available in the MicroLab
PC software:
Data Collect: Only collect and display the infrared spectrum. No
further analysis is performed on the collected data.
Component: Quantitative prediction methods. An infrared
spectrum is collected then a predefined quantitative prediction is
made from that data. Several components can be defined, each
with their own quantitative calibration.
Qualitative Library Search: Collect an infrared spectrum and
search it against a predefined spectral library. The results of this
method will be the top matches to the spectral library.
Check for Clean Crystal Prior to Collecting Background: This
option enables or disables the system checks of the cleanliness of the
ATR crystal or sample cell prior to collecting a background. This
option is recommended.
Background: Two options exist for collection of the background:
New Background for Each Sample: This requires the collection
of a new background spectrum for every sample. This is the
recommended option, as it provides the best correction for water
vapor, carbon dioxide and other atmospheric variations.
Collect Background at Time Interval: If this option is selected,
samples can be collected in sequence without measuring a
background spectrum in between samples.