Technical data
MicroLab Mobile
14 Agilent MicroLab Software Operation Manual
Start
The Start button begins a sample measurement using the currently
selected method. The currently selected method is shown in the
upper left corner of the Home screen.
Method
The Method button opens the Method Selection dialog box. Users
without Admin privileges can only select methods designated as
Public by the administrator. See the Method Selection instructions
below.
Reference Method
The Reference Method button enables the collection of a reference
sample for a component method. Some component methods present
concentrations as a percentage increase or decrease. For these
calculations, a reference of the in specification sample is needed to
calculate these starting values. Also, the Reference Method saves the
spectrum of the in specification (new) material for later visual
comparison.
Log Off
The Log Off button exits the current user and returns the MicroLab
Mobile software to the Logon screen. From there, the Exit button can
be tapped to close the software.
Menu
The menu contains:
Previous Results
Advanced Features
Previous Results
Enables viewing of previously collected data. Only Service, Admin
and Developer roles have access to Previous Results.