Data Sheet
CX  Auto-Calibrate Crosstalk  0x43 0x58…  
Perform Crosstalk Calibration on tinyLiDAR. This correction factor is typically 
not used as it is meant to try and correct for crosstalk from thin cover 
windows placed in front of the sensor. The tinyLiDAR
 module does not ship 
with a cover window hence it does not require crosstalk correction by 
default.  
Before using this calibration command, please set tinyLiDAR to 
Continuous, 
High Accuracy
 mode by issuing the commands “MC” and “PH”. 
ST recommends to use a distance of 400mm to a gray (17% reflectivity) 
target. Therefore, place a gray target 400mm away from tinyLiDAR
 before 
running this calibration.  
Must specify calibration distance in mm.  
The crosstalk factor will be placed in non-volatile memory and used for all 
subsequent operations. This calibration takes about 10seconds to run and 
the blue LED will flash slowly during the calibration. You can clear this 
calibration out by running the “RESET” command. 
Before using this calibration command, please set tinyLiDAR to 
High 
Accuracy
, 
Continuous 
mode by issuing the commands “PH” and “MC”. 
Host -> Sensor 
Send Command 
0x43 {C}, 0x58 {X}, Distance to Cal Target in mm 
Example:  
Run Crosstalk calibration for the tinyLiDAR connected at the default address 
of 0x10 with a target exactly 400mm away:  
<10:w> 43 58 01 90  
tinyLiDAR Reference Manual rev1.26  20 