Instruction manual

Troubleshooting
continued
RFT9739 Rack-Mount Transmitter Instruction Manual
75
Before You Begin Getting Started Mounting
Power-Supply and
Sensor Wiring
Output Wiring Startup Troubleshooting
Table 7-3. Using overrange and sensor error messages
Table 7-4. Using slug flow and output saturated messages
Instructions
1. Turn off power to transmitter.
2. Unplug terminal blocks from transmitter back panel to check circuits.
Message Other symptoms Cause(s) Corrective action(s)
Drive Overrng or
Input Overrange
Transmitter produces fault outputs
At the transmitter, DMM indicates open or
short circuit from red wire to brown wire
At the sensor, DMM indicates open or short
circuit from red wire to brown wire
Flow rate outside
sensor limit
Faulty cable
Open or short drive
coil in sensor
Fill sensor with process fluid
Bring flow rate within sensor
limit
Monitor flow rate
If open or short at
transmitter, reconnect wiring
or repair cable
If open or short at sensor,
return sensor to Micro
Motion
Transmitter produces fault outputs
At the transmitter, DMM indicates open or
short circuit from green wire to white wire
At the sensor, DMM indicates open or short
circuit from green wire to white wire
Flow rate outside
sensor limit
Faulty cable
Open or short left
pickoff in sensor
Sensor Error Transmitter produces fault outputs
At the transmitter, DMM indicates open or
short circuit from blue wire to gray wire
At the sensor, DMM indicates open or short
circuit from blue wire to gray wire
Faulty cable
Open or short right
pickoff in sensor
If open or short at
transmitter, reconnect wiring
or repair cable
If open or short at sensor,
return sensor to Micro
Motion
Transmitter produces fault outputs Moisture in sensor case Replace conduit and/or
conduit seals
Repair cable
Return sensor to Micro
Motion
Drive Overrng or
Dens Overrng
Transmitter produces fault outputs Inappropriate density
factors
Process density
> 5.0000 g/cc
Severely erratic or
complete cessation of
flow tube vibration due
to gas slugs or solids
in process fluid
Plugged flow tube
Calibrate for density
Correct density factors
Monitor density
Bring density within sensor
limit
Purge flow tubes with
steam, water, or purging
chemical
Temp Overrange Transmitter produces fault outputs
At the transmitter, DMM indicates open or
short circuit from yellow wire to orange wire
At the sensor, DMM indicates open or short
circuit from yellow wire to orange wire
Temperature outside
sensor limit
Faulty cable
Open or short lead
length compensator
Bring temperature within
sensor limit
Monitor temperature
If open or short at
transmitter, reconnect wiring
or repair cable
If open or short at sensor,
return sensor to Micro
Motion
Transmitter produces fault outputs
At the transmitter, DMM indicates open or
short circuit from violet wire to yellow wire
At the sensor, DMM indicates open or short
circuit from violet wire to yellow wire
Faulty cable
Open or short RTD in
sensor
Message Condition Corrective action(s)
Slug flow Gas slugs causing process density to go below low slug flow limit
Solids causing process density to go above high slug flow limit
Monitor density
Enter new slug flow limits
Enter new slug duration
Freq overrange Flow rate driving output from terminals CN2-D24 (FREQ) and
CN2-D26 (RETURN) to 0 or 15 kHz
Change flow measurement units
Rescale frequency/pulse output
Reduce flow rate
mA 1 saturated Output from terminals CN2-Z30 (PV+) and CN2-D30
(PV–) equals 0, 3.8, or 20.5 mA
Change value of variable at 20 mA
Alter fluid process
mA 2 saturated Output from terminals CN2-Z28 (SV+) and CN2-D28
(SV–) equals 0, 3.8, or 20.5 mA