RFT9739 Field-Mount Transmitter Instruction Manual
19
Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mounting
Power-Supply and
Sensor Wiring
Figure 4-3.
Power-supply wiring
terminals
Grounding
To ensure proper grounding:
• If the installation must comply with UL, CSA, or SAA standards, refer to
the instructions in one of the following Micro Motion documents:
-
UL-D-IS Installation Instructions
-
CSA-D-IS Installation Instructions
-
SAA-D-IS Installation Instructions
• In most installations, install grounding as illustrated in
Figure 4-4a
(next page).
• For installations in Europe, and to comply with CENELEC standards,
install grounding as illustrated in
Figure 4-4b
(next page).
• To achieve potential equalization and comply with CENELEC
standards for hazardous area installations in Europe, the RFT9739
external ground terminal (see
Figure 4-4b
) should be connected to the
appropriate ground terminals within the hazardous area, using a
potential equalizing line.
Equipment
ground
Power-supply conduit opening
Equipment
ground
Transmitter
power rating
Terminal
labels
85-250 VAC 50/60 Hz L N
12-30 VDC +
WARNING
Failure to comply with requirements for intrinsic
safety in a hazardous area could result in an
explosion.
The transmitter must be properly grounded.
Follow the instructions below to ground the transmitter.
For hazardous area installations in Europe, refer to
standard EN 60079-14 if national standards do not apply.