Figure G-3. RFT9712 Remote Flow Transmitter terminals
Table G-3. RFT9712 to RFT9739 terminal conversions
Take the wire from
RFT9712 terminal
number:
...and connect it to
RFT9739 terminal
number:
Wire color
(Micro Motion
color-coded cable) Function
00 Black
[1]
[1]
Combined shields from brown/red, green/white, and gray/blue pairs
Shields
11 Brown Drive +
22 Red Drive –
3 3 Orange Temperature –
4 4 Yellow
[2]
[2]
Shield wire from orange/violet pair.
Shield (Temperature lead length
compensation)
5 5 Green Left pickoff +
6 6 Blue Right pickoff +
7 7 Violet Temperature +
8 8 Gray Right pickoff –
9 9 White Left pickoff –
10
11
12
See RFT9739 power-supply wiring
and grounding instructions (
Chapter 4
)
14 16 — Return
15 21 — Zero +
16 18 — PV –
17 17 — PV +
18 16 — Return
19 15 — Freq +
21 27 — 485A
22 26 — 485B
23 23 — Signal ground
24 24 — Temperature
25 25 — Tube period
26 22 — Control