Troubleshooting30 Q3 Dosing Module
Error symptoms Possible cause Diagnostic Remedy
The value on the display is
unstable when dosing
powders.
The weighing sample is
electrostatically charged.
Check if the weighing
result is stable when using
a test weight.
Increase the air humidity
in the weighing chamber.
Use an antistatic kit. See
"Accessories" in the
Reference Manual (RM).