Reference Manual
Table Of Contents
- Title Page
- Table of Contents
- English
- Table of Contents
- 1 Introduction
- 2 Safety Information
- 3 Design and Function
- 4 Installation and Putting into Operation
- 5 Operation
- 6 Maintenance
- 7 Troubleshooting
- 8 Technical Data
- 9 Disposal
- 10 Accessories and Spare Parts
- 11 Appendix
- Index
Draft
Maintenance28 Q3 Dosing Module
§
The liquid dosing head is installed on the dosing or liquid module.
§
A sample vessel is on the weighing pan, big enough for the quantity of liquid in the liquid tube.
1 Empty the liquid tube by replacing the bottle by an
empty bottle and using the Purge function.
ð The liquid tube is empty.
2 Release pressure.
3 Remove the dosing head from the dosing or liquid
module.
4
WARNING: Injury or contamination from
hazardous substances. Be aware of any liquid
draining from the tubing and the valve.
Open the dosing head with the micro dosing valve tool.
5 Take out the micro dosing valve and clean it, e.g., in an ultrasonic bath.
6 After the cleaning, reinstall the micro dosing valve and refill/replace the bottle.
6.2 Replacing the sealing ring and fastening nut on the liquid tube
CAUTION
Injury due to splashing liquids
If the pressure in the bottle is not released, liquid might splash when removing the micro
dosing valve, opening the bottle, or removing the liquid tube.
− Always release pressure before removing the micro dosing valve, opening the bottle, or
removing the liquid tube.
Removing the sealing ring and the fastening nut
If liquid is leaking from the dosing head nut, replace the sealing
ring and fastening nut at the dosing head (1). If pressure cannot
be built in the bottle, replace the sealing ring and fastening nut at
the bottle cap (2).
1
2
1
2
§
Pressure is released.
1 Unscrew the fastening nut at the dosing head or the
bottle cap.
2 To access the sealing ring (1), slide back the fastening
nut (2).
3 If changing the nut and ring at the bottle cap, remove
the suction filter and slide the sealing ring all the way
to the bottle-end of the tube.
4 Cut off the tube above sealing ring (1). Check that the
end of the tube is straight.
5 Remove the fastening nut.
Reconnecting the liquid tube
Insert the new sealing ring and fastening nut and reconnect the liquid tube.
See also