Operation 23Q3 Dosing Module
5.4 Liquid dosing
5.4.1 Attaching the dosing head to the bottle cap
1 Insert the liquid dosing head in the liquid dosing head
support.
2 To remove the liquid dosing head from the liquid dosing head
support, pull the lever (1) towards the dosing head and
remove the liquid dosing head (2).
1
2
5.4.2 Handling the bottle
CAUTION
Injury due to splashing liquids
If the pressure in the bottle is not released, liquid might splash when removing the micro
dosing valve, opening the bottle, or removing the liquid tube.
− Always release pressure before removing the micro dosing valve, opening the bottle, or
removing the liquid tube.
Filling the bottle
The suction filter (1) has to be covered with liquid at all times.
Before the suction filter gets dry, refill the bottle.