Operation20 Q3 Dosing Module
5 Operation
For further information, consult the XPR Reference Manual (RM).
u www.mt.com/XPR-analytical-RM
5.1 Installing and removing the dosing head
When the dosing module is in its upper-most position, the dosing head is unlocked. That means that it can be
released by pressing the release buttons of the holder. During dosing, the dosing head is locked and cannot be
removed until it returns to its upper-most position.
Installing the dosing head
1
2
§
The dosing module is in the upper-most position.
1 Slide the dosing head (1) onto the dosing head holder
(2) until it comes to a stop.
2 Slightly press down the dosing head until it properly
sits in the dosing head holder (2).
ð The dosing head is locked in the holder. Check that
it is secured by slightly pulling on it.
ð The dosing head is ready to dose.
Removing the dosing head
1
2
§
The dosing module is in the upper-most position.
1 Press the release button (1) on either side of the
dosing head holder to release the dosing head (2).
2 Simultaneously, pull the dosing head (2) up and
outwards.