Installation and Putting into Operation16 Q3 Dosing Module
6 Using the micro dosing valve tool, slide the sealing
ring until the distance (D) is enough for the tube to
reach the bottom of the bottle. Recommended
distances for typical bottle volumes are listed next.
Typical distances (D) between the sealing ring and the
bottle-end of the tube
Thread Bottle volume Distance (D)
GL45 1000ml 220mm
500ml 170mm
250ml 135mm
GL25 25ml 90mm
Connecting the liquid tube
4
1
2
3
5
1 Insert the dosing-head end of the tube (1) in the
dosing head (2).
2 Tightly fasten the fastening nut to the dosing head.
3 Insert the bottle end of the tube through the corre-
sponding hole in the bottle cap (3). The tube should
reach the bottom of the bottle.
4 If needed, attach the suction filter (4) to the bottle end
of the tube.
5 Tightly fasten the fastening nut to the bottle cap.
6 Screw the cap to the bottle (5).
The suction filter is used to ensure that no particles or
impurities are carried through the liquid dosing head.
Using the suction filter will extend the life time of the
dosing head. However, when dosing solutions, molecules
of one of the substances might be absorbed by the suction
filter, altering the concentration of the solution. The suction
filter should only be used when dosing pure solvents.
Connecting the air tube
WARNING
Injury and/or damage due to reacting substances
When pressure is released from the bottle, the air/gas in the bottle moves back towards the
pump. The air/gas coming from the coupled outlets mixes in the pump. Molecules of the
substances in the various bottles can get in contact through this contaminated air/gas.
1 Do not connect bottles with incompatible liquids to the same pump simultaneously.
2 Before connecting a second, incompatible liquid to the pump, disconnect the first bottle
and purge the pump cavity with clean air/gas.