Installation and Putting into Operation 15Q3 Dosing Module
4 Insert the plug of the power cable into a grounded power outlet that is easily accessible.
5 Using the USB cable, connect the USB port of the pump to the USB device port of the balance.
4.3.4 Connecting the tubes
Tubes definition
The liquid tube is the thinner tube (1) used for transporting liquid
from the bottle to the liquid dosing head. The air tube is the
slightly bigger tube (2) used for pumping air into the bottle. By
adding air through the air tube, pressure rises in the bottle. When
the pressure reaches the configured target pressure, between 0.3
and 0.5bar (4.4 to 7.2psi), the micro dispensing valve in the
dosing head opens and liquid can ascend the liquid tube. .
1
2
1Liquid tube
2Air tube
Preparing the liquid tube
CAUTION
Injury due to leaking liquids
Wrongly cut tubing can result in leaking connections.
Cut the tubes with a tube cutter or a sharp knife.
§
The liquid dosing head is inserted in the liquid dosing
head support of the bottle.
1 Using a tube cutter or a sharp knife, cut a sufficient
amount of tubing. The appropriate length mainly
depends on the distance between the bottle and the
dosing module while dosing.
Recommended length: about 0.9m
0.9 m
D
1 2 2 3
2 Place the sealing ring (1) on a flat stable surface, e.g.,
a table or a workbench, with the wider end
downwards.
3 Take the end of the liquid tube and press it into the
sealing ring.
ð This is the dosing-head end of the tube. The
opposite end is the bottle end.
4 Thread the fastening nuts (2), paying attention to the
orientation.
5 Thread the sealing ring (3) from the bottle end of the tube. Pay attention to the orientation.
en