Reference Manual

Table Of Contents
Draft
Installation and Putting into Operation 13Q3 Dosing Module
4 Installation and Putting into Operation
This instrument must be installed by a METTLER TOLEDO service technician.
4.1 Selecting the location
A balance is a sensitive precision instrument. The location where it is placed will have a profound effect on the
accuracy of the weighing results.
Requirements of the location
Place indoors on stable
table
Ensure sufficient spacing Level the instrument Provide adequate lighting
Avoid direct sunlight Avoid vibrations Avoid strong drafts Avoid temperature fluctu-
ations
Sufficient spacing for balances: > 15 cm all around the instrument
Take into account the environmental conditions. See "Technical Data."
Note
The powder delivery mechanism of the Q3 dosing module might cause the instrument to vibrate. Do not place
on the same working surface as other instruments that are sensitive to vibrations.
4.2 Scope of delivery
Q3 dosing module
Dosing module
Dosing lift
Top door dosing module
Sealing inserts, 5 pcs
Powder dosing head
ErgoClip adapter holder
Various vial adapters, 4 pcs
Declaration of conformity
User Manual
Recommended options
EasyHub USB Antistatic kit
The scopes of delivery for the QL3 pump and the QLL kit for bottles are in the documentation delivered with
these products.
en