Reference Manual
Table Of Contents
- Title Page
- Table of Contents
- English
- Table of Contents
- 1 Introduction
- 2 Safety Information
- 3 Design and Function
- 4 Installation and Putting into Operation
- 5 Operation
- 6 Maintenance
- 7 Troubleshooting
- 8 Technical Data
- 9 Disposal
- 10 Accessories and Spare Parts
- 11 Appendix
- Index
Draft
Design and Function 11Q3 Dosing Module
Optical sensor for vial detection
The dosing module is equipped with an optical sensor
to detect the position of the vial opening. This allows
automatic vertical positioning of the dosing module
for fast and safe dosing.
ErgoClip adapter holder and adapters
The Q3 dosing module includes an adapter holder
that can be placed on the SmartGrid weighing pan. It
can hold adapters of various sizes allowing perfect
positioning of different sample vessels. Additional
adapters are available as accessories.
See also
2 Accessories}Page42
3.4 Dosing heads and equipment
Powder dosing head
Powder dosing heads are used for automatic powder dosing. If
only a small quantity of powder is required, the dosing head can
be used without a vial attached to it. In this case, the opening
can then be closed using the spare cap included in the delivery.
The dosing heads can be stored in appropriate storage
containers. Vials and storage containers need to be purchased
separately.
If a printer is connected, you can use a pre-defined template to
print a label with the dosing-head data from the RFID tag and
affix this label to the dosing head.
Powder test head
The powder test head is used to test a powder dosing system.
The dosing head is filled with calcium carbonate (CaCO
3
) and
uses a predefined test program to check the dosing functionality
of the system. The powder test head is programmed to perform
15 tests. Each test weighs 10 doses of a pre-defined amount of
powder. If your instrument fails the tests, inform a METTLER
TOLEDO service technician.
en