Operating instructions

Further important information regarding your PG-S balance
53
Error message
Rbort \
Balance blocked
Cause
Adjustment (calibration) or test could not be
performed properly.
The balance aborts the procedure. This error
message is caused by external disturbing
influences (e.g. vibrations or powerful
drafts).
In the menu option Handshake, setting
V
HS
Hard" selected.
Rectification
Press «C» key (a double beep sounds as
confirmation) to clear the error message.
Close all draft shield doors.
Possibly select a more suitable location for
the balance.
Switch on attached peripheral device.
6.2 Preventive maintenance and care
Servicing
Regular servicing of your balance by an authorized service engineer ensures constant accuracy for years to come and
prolongs the lifetime of the instrument. Ask your METTLER TOLEDO dealer for details of the available service options.
Cleaning
The balance housing and the weighing pan are made of high-grade, resistant
materials. All commercially available cleaning agents may thus be used for
cleaning.
PG-S balances can best be cleaned with a damp cloth.