Operating instructions

Further important information regarding your PG-S balance
52
6 Further important information
6.1 Error messages
Error messages in the display draw your attention to incorrect operation or that the balance could not execute a procedure
properly.
Error message
t 1 \
\
\
\
F nonE \
Error I \
\
Error 3 \
\
\ 1 / \
nnn \
~ uuu ~ \
/ 1 \ \
Cause
Overload
Underload
No function preselected
No stability
- On taring or adjustment (calibration)
- On loading the reference weight for the
functions "Piece counting" or "Percent
weighing".
Faulty reference
(Reference weight or reference piece number
too low)
Wrong or missing weighing pan
Rectification
Unload weighing pan
Check that the weighing pan is positioned
correctly.
Preselect desired function in the menu.
Ensure more stable ambient conditions. If
not possible, check settings of the repeat-
ability and vibration adapter (see Sections
4.9 and 4.7).
Increase reference weight or reference piece
number.
Mount correct weighing pan.
Unload weighing pan.