Operating instructions
Special applications and functions
48
Place the requested weight in the middle of the weighing pan.
Cal
\ 1
nnn
~ U.UU"
/ 1
CRL don
flborfc
\
A
\
- BALANCE CALIBRATION -
03.10.97 11:34:23
METTLER TOLEDO
Type: PG2002-S
SNR: 1105238536
SW: 1.74 1.3
Weight ID:
Weight: 1000.00 g
External cal. done
Signature:
END
During the adjustment the horizontal segments are displayed.
Note
You can abort the ongoing adjustment at any time by pressing the «C» key briefly.
On completion of the adjustment operation, you are prompted to remove the
weight. Lift the weight off the weighing pan.
After removal of the weight, the balance shows the end of the adjustment
operation and then returns to the weighing mode.
Note
If the adjustment (calibration) can not be performed properly (e.g. as a result of
vibrations), the balance aborts the adjustment operation and "Abort" appears in
the display. Press the «C» key to clear this message and restart the adjustment
operation.
If your balance is connected to a printer, the adjustment is recorded automatically
in conformance with GLP. The record shown opposite is a specimen printed with
the METTLER TOLEDO LC-P45 Printer. Depending on the attached printer, the
printout may differ somewhat from the example shown.