Operating instructions
Special applications and functions
46
The following displays appear during the adjustment (calibration):
Cal
The internal adjustment weight is being loaded.
Cal
nnn
u.uu
Cal
g
Rbork
The internal adjustment weight is being raised.
The balance is processing the adjustment results.
- BALANCE CALIBRATION -
03.10.97
11:23:34
METTLER TOLEDO
Type:
SNR:
SW:
Internal cal
Signature:
PG802-S
1105238536
1.74 1.30
. done
n
The balance reports successful completion of the adjustment (calibration).
The balance automatically returns to the weighing mode.
You can always abort an ongoing adjustment (calibration) by briefly pressing
the «C» key (double beep).
If the adjustment (calibration) can not be performed properly (e.g. as a result of
vibrations), the balance aborts the adjustment operation and "Abort" appears in
the display. Press the «C» key to clear this message and restart the adjustment
operation.
If your balance is connected to a printer, the adjustment (calibration) is recorded
automatically in conformance with GLP. The record shown opposite is a
specimen printed with the METTLER TOLEDO LC-P45 Printer. Depending on the
attached printer, the printout may differ somewhat from the example shown.