Operating instructions
Special applications and functions
45
5.6 Adjustment (calibration) with internal weight
Depending on the setting selected in the menu (see Section 4.4), the adjustment (calibration) can be performed with the
built-in, internal weight fully automatically (FACT) orsemi-automatically.
Fully automatic internal adjustment (calibration) FACT
Your balance is set in the factory for the fully automatic adjustment with the
internal adjustment weight. You are already familiar with this setting from
Sections 2.5 and 4.4.
Semi-automatic adjustment (calibration)
If your balance is outside the adjustment tolerance and depending on whether
you have set the automatic adjustment call-up in the menu (see Section 4.5),
the balance uses a flashing «Cal» in the display to prompt you to adjust
(calibrate) with the internal weight at a keystroke. With certified balances, the
adjustment (calibration) with the internal weight is performed automatically in
accordance with the national weights and measures legislation. An adjustment
(calibration) with external weights is not allowed by the weights and measures
legislation.
If you wish to adjust your balance with an internal weight, proceed as follows:
Make sure that "FACT" or the "Adjustment (calibration) with internal weight
(Cal int)" is selected in the menu (see Section 4.4).
Ensure that the weighing pan is unloaded and close the doors of the draft shield
(if used). There is no need to tare the balance before the adjustment (calibration).
l/io d
Cal
:RL mt
Start the adjustment operation by pressing and holding the «Cal» key. The
balance briefly shows that adjustment (calibration) is being performed with the
internal weight.
long Note
If "SECUrEd 2" is switched on in the menu, the ™< key is blocked.