Operating instructions

Special applications and functions
42
5.4 Dynamic weighing of unstable weighing samples
The functions "Dynamic weighing with automatic start" and "Dynamic weighing with manual start" facilitate your weighing
of unstable weighing samples (e.g. animals). With this type of weighing, your balance determines the weight over a
particular time period and calculates a representative mean value.
Dynamicweighing presupposes thatyou have preselected the function "Fdyn A" or "FdynM" inthe menu (see Section
4.6).
If you are working with a weighing container, place it on the weighing pan in the
normal weighing mode.
f
/
s
B
IHHl
U.UU
nnn
U.UU
A
\
9 \
\
9 \
Press the «->0/T<-» key to tare the balance.
Press the «!£3» key briefly. The symbol of the weighing process adapter in the
display confirms that dynamic weighing has been activated.
Your balance is set in the factory so that the weight is determined over a period
of 3 seconds. You need perform the following 3 steps only if you wish to change
this time interval.
H
t r -3
"
Press and hold the «F» key until the time display appears.
long