The menu
28
4.12 Switching automatic zero correction (Autozero) on or off
This menu option allows you to switch the automatic zero correction on or off. When it is switched on (factory setting),
the zero point is automatically corrected for drift or contamination of the weighing pan.
The following settings are available:
He!
on
Autozero switched on
This is the factory setting. The zero point is automatically corrected.
R?. off
Autozero switched off
The zero point is not automatically corrected. This setting is advantageous for
special applications (e.g. evaporation measurements).
Note
With certified balances, this setting is possible only with a resolution of e= 10 d.
4.13 Preselecting automatic shutdown
When the automatic shutdown is activated, the balance automatically switches itself off after a preselected time (calculated
from the last operation) and is switched to the standby mode.
The following settings are available:
RoFF
No automatic shutdown
The automatic shutdown is deactivated (factory setting).
RoFF
Automatic shutdown after 2 minutes
As soon as the balance has not been operated for 2 minutes, it switches itself off
automatically.